Advantest R3261C

Spectrum Analyzer, 9kHz-2.6GHz
Features
R3261C,
R3261CN, R3261D, R3361C, R3361CN, R3361D
- Total
Level Accuracy of 1 dB (Typical)
- User-defined
Functions
- Measuring
Window Function
- 120 dB
Display Range
- Built-in
Tracking Generator (R3361C, R3361CN, R3361D)
- CDMA
Power Measurements (with Option 1K)
R3261/R3361
Series Spectrum Analyzers
The
R3261/R3361 Series spectrum analyzers use a synthesized technique to cover
wide frequency bands: 9 kHz to 2.6 GHz (R3261C, R3361C) or 9 kHz
to 3.6 GHz (R3261D, R3361D). These compact analyzers also offer high-performance
functions, such as the 1 Hz resolution frequency setting function
and 1 Hz resolution frequency counter function.
The
R3261/R3361 Series spectrum analyzers are easy to operate because they
incorporate unique features such as user-defined and measuring window
functions. Weighing only 15 kg (33 lbs.) the analyzers are small and useful
for measurement and analysis of all kinds, from manufacturing to research
and development.
The
analyzers have an internal controller function (Option 15), parallel I/0,
and GPIB interface for line connection and automatic measurement, so the
user can create the optimum system configuration.
Choose
From Six Models, Depending Upon Your Application
The
R3261/R3361 Series consists of six models. All models feature high performance
and a set of features and functions for various applications.
|
|
R3261C
|
R3261CN
|
R3261D
|
R3361C
|
R3361CN
|
R3361D
|
|
Frequency
Range
|
9 kHz
to
2.6 GHz
|
9 kHz
to
2.6 GHz
|
9 kHz
to
3.6 GHz
|
9 kHz
to
2.6 GHz
|
9 kHz
to
2.6 GHz
|
9 kHz
to ,br>3.6 GHz
|
|
Input
Impedance
|
50 Ohm
|
75 Ohm
|
50 Ohm
|
50 Ohm
|
75 Ohm
|
50 Ohm
|
|
Internal
Tracking
Generator
|
Not
available
|
Not
available
|
Not
available
|
Standard
|
Standard
|
Standard
|
|
Memory
Card
|
Standard
|
Standard
|
Standard
|
Standard
|
Standard
|
Standard
|
|
Controller
Function
|
Opt.
15
|
Opt.
15
|
Opt.
15
|
Opt.
15
|
Opt.
15
|
Opt.
15
|
|
Occupied
Bandwidth
Measurement,
Adjacent-channel
Leakage,
Power Measurement
|
Standard
|
Standard
|
Standard
|
Standard
|
Standard
|
Standard
|
|
Burst
Signal Analysis
|
Opt.
12
|
Opt.
12
|
Opt.
12
|
Opt.
12
|
Opt.
12
|
Opt.
12
|
|
CDMA
Power
Measurement
|
Opt.
1K
|
N/A
|
Opt.
1K
|
Opt.
1K
|
N/A
|
Opt.
1K
|
1 Hz
Resolution Synthesizer
The
R3261/R3361 Series portable spectrum analyzers are based upon a synthesized
system, so the center and start/stop frequencies can be set with a resolution
of 1 Hz. The synthesizer, featuring 1 Hz resolution, is a powerful
tool ideal for waiting, receiving, or spot measurement in the zero-span
mode.
1 Hz
Resolution Frequency Counter
The
frequency counter built into the R3261/R3361 Series features 1 Hz
resolution. One of the advantages of these analyzers is that they can
measure the modulated frequency or spurious frequency of a radio system
that cannot be measured with an ordinary frequency counter simply by setting
a marker. In addition, the ability to measure weaker signals than a frequency
counter can extend the application range from broadband panoramic measurement
to weak signal measurement.
Manual
Sweep Function for Spot Measurement
EMC
measurements using the quasi-peak (QP) detector require an extended sweep
time. However, the time required to measure the peak value can be shortened
using the manual sweep function.
Create
Your Own Unique Menu
The
R3261/R3361 Series spectrum analyzers are very functional and very easy
to use because of the user-defined functions:
- User-defined
function. The user can define a necessary function on the USER
key in the same way as on the function keys of a personal computer.
A unique menu can be created.
- Define
function. The define function enables the user to change the software
key menu manually. With this function, you can create a unique system
by changing the key functions that were defined before the system
was shipped from the factory.
Fast
Measurement with Measuring Window Function
The
R3261/R3361 Series allows the user to define both frequency and level
ranges within a span for analysis. This powerful tool enables faster measurement
times over frequencies and levels of interest.
Option
1K includes a suite of CDMA test routines which can be accessed through
softkey menus. Specific CDMA measurements include channel power, total
power across the span, average power, and peak power. The softkey menu
for a test along with measured data (waveforms) can be stored in memory
cards via the front panel.
1 kHz to 2.6 GHz and zero.
R3261D/R3361D, 1 kHz to 3.6 GHz and zero.
Log mode: 1, 2, or 3 decades selected between 10 kHz and 1000 MHz.
Frequency Span Accuracy -
Span <2 MHz: </=±3% of span.
Span >2 MHz: </=±5% of span.
Frequency Stability - Residual
FM:
10 MHz </= span < 2 MHz, 50 kHz p-p or less.
Span >2 MHz, 2 kHz p-p or less.
Span </=2 MHz, 20 Hz p-p or less.
Frequency drift: Span </=2 MHz, 300 Hz/min. or less.
Sideband Noise - At 20 kHz
offset
f </= 3.0 GHz, -105 dBc/Hz;
f </= 3.6 GHz, -101 dBc/Hz.
Resolution -
3 dB bandwidth: 30 Hz to 1 MHz; switchable in 1 to 3 steps.
6 dB bandwidth: 200 Hz, 9 kHz, 120 kHz. Selectivity:
</=15:1 (60 dB: 3 dB).
Bandwidth accuracy: </=20%.
Marker Accuracy -
Normal mode: Central frequency display accuracy + span accuracy.
Counter mode: Display frequency x reference oscillator accuracy ±1
count.
Amplitude Related
Amplitude Measuring Range -
R3261C, R3261D, R3361C, R3361D: -130 dBm to +25 dBm.
R3261CN, R3361CN: -19 dBµV to +132 dBµV.
Screen Display Range -
Log mode: 10 dB/div, 120 dB. 10 dB/div, 80 dB. 5 dB/div,
50 dB. 2 dB/div, 20 dB. 1 dB/div: 10 dB.
Linear mode: 10 div.
QP mode: 80 dB (10 dB/div) when measuring range is 70 dB.
Display Linearity -
Log mode: ±2.0 dB/110 dB, ±1.5 dB/70 dB,
±1.0 dB/10 dB, ±0.2 dB/1 dB.
Linear mode: ±5% of full scale.
QP mode: ±2.0 dB/70 dB, ±1.0 dB/40 dB.
Reference Level Display Range
-
R3261C, R3261D, R3361C, R3361D:
-109.9 dBm to +40.0 dBm, 0.715 µV to 22.4 V. R3261CN,
R3361CN: +0.1 dBµV to
+150 dBµV, 1.01 µV to 31.6 V.
Reference Level Accuracy (after Automatic
Calibration) -
|
|
R3261C/R3261D
R3361C/R3361D
|
R3261CN
R3361CN
|
|
<±0.3 dB
|
0 dBm
to -50 dBm
|
+110 dBµV
to +60 dBµV
|
|
<±0.7 dB
|
+20 dBm
to -70 dBm
|
+130 dBµV
to +40 dBµV
|
Dynamic Range
Average Noise Level - (resolution
bandwidth - 300 Hz, Video bandwidth 1 Hz, Input attenuator
0 dB, Frequency range 10 MHz or more):
R3261C, R3261D, R3361C, R3361D:
-120 dBm + 1.55 f (GHz) dB. R3261CN, R3361CN: -10 dBµV
+1.55 f (GHz) dB.
Secondary and Tertiary Distortion
-
</=-70 dB at -30 dBm input, Input attenuator 0 dB, Frequency
10 MHz or more.
Frequency Response -
|
|
R3261C/R3261D
R3361C/R3361D
|
R3261CN
R3361CN
|
|
100 Hz to 2 GHz
|
</=+0.5 dB
|
</=+0.5 dB
|
|
9 kHz to 2.6 GHz
|
</=+1.0 dB
|
|
|
9 kHz to 3.6 GHz
|
|
</=+1.5 dB
|
Residual Response -
R3261C, R3261D, R3361C, R3361D:
</=-100 dB (Termination: 50 Ohm). R3261CN, R3361CN: </=-11 dBµV
(Termination 75 Ohm).
Resolution Bandwidth Switching Accuracy - </=±0.3 dB after
automatic calibration.
Video Filter - 1 Hz to 1 MHz; switchable in 1 or 10 steps.
Sweep Related
Sweep Time - 50 ms to 1000
s and manual sweep.
Sweep Time Accuracy - </=3%.
Trigger Modes - FREE RUN,
LINE, VIDEO, EXT, TV-V, and SINGLE.
Tracking Generator Specifications (R3361C, R3361CN, R3361D)
Frequency Range - R3361C/R3361CN:
9 kHz to 2.6 GHz. R3361D: 9 kHz to 3.6 GHz.
Output Level Range - R3361C,
R3361D: 0 dBm to -50 dBm. R3361CN: +105 dBµV to 55 dBµV
setting in 1 dB steps.
Output Level Accuracy - </=±0.5 dB
(30 MHz, -10 dBm, 20°C to 3O°C).
|
|
R3361C/R3361D
(at -10 dBm
output)
|
R3361CN
(at +95 dBµV
output)
|
|
100 kHz to 1.0 GHz
|
</=0.7 dB
|
</=0.7 dB
|
|
9 kHz to 2.6 GHz
|
</=1.5 dB
|
</=1.5 dB
|
|
9 kHz to 3.6 GHz
|
</=2.0 dB
|
</=2.0 dB
|
Output Level Switching Accuracy
(at 0 dBm/+95 dBµV output) -
100 kHz to 1.0 GHz: </=±1.0 dB.
9 kHz to 2.6 GHz: </=±2.0 dB.
9 kHz to 3.6 GHz: </=±3.0 dB.
Output Spurious (at 0 dBm/+105 dBµV
output) -
Harmonic spurious: </=-20 dB.
Non-harmonic spurious: </=30 dB.
Tracking Generator Leakage -
R3361C/R3361D: Frequency </=3.0 GHz,
</=-110 dBm. Frequency </=3.6 GHz,
</=-100 dBm.
R3361CN: </=+1 dBµV.
Output Impedance -
R3361C, R3361D: Approximately 50 Ohm.
R3361CN: Approximately 75 Ohm.
Output VSWR (at </=-10 dBm/+95 dBµV
output) - 100 Hz to 2 GHz: </=1.5. 9 kHz to 3.6 GHz:
</=2.0.
Output Connector - N type.
Input Related
Input Impedance -
R3261C, R3261D, R3361C, R3361D: 50 Ohm.
R3261CN, R3361CN: 75 Ohm.
VSWR (at input attenuator >/=10 dB)
-
100 kHz to 2 GHz: </=1.5.
9 kHz to 3.6 GHz: </=2.0.
Input Connector - N type.
Maximum Input Level -
R3261C, R3261D, R3361C, R3361D:
+25 dBm (attenuator >/=30 dB), 50 V DC max. R3261CN,
R3361CN: +132 dBuV (attenuator >/=30 dB), ±50 V
DC max.
Input Attenuator - 0 to 50 dB
in
10 dB steps.
Input Attenuator Switching Accuracy
- </=1.0 dB (</=2.0 GHz), </=1.5 dB (</=3.6
GHz) at input attenuator 10 dB.
Detection Modes - NORMAL,
POSI, NEGA, and SAMPLE.
Inputs/Outputs
External Memory Function - IC
memory card.
GPIB Data Output/Remote Control
- Data output and remote control through internal GPIB interface.
Direct Plotting - Hardcopy
output of all display data to HPGL plotters through internal GPIB interface.
Voice Monitor Output - AM
and FM with approximately 8 Ohm earphone.
Probe Power Source: - ±15 V,
4-Pin connector.
Environmental
Ambient Temperature - 0°C
to +50°C. Relative Humidity: 85% or less.
Storage Temperature - -20°C
to +60°C.
Safety - This product has
been safety tested by Advantest to IEC 348 specifications.
Power Requirements
Operating Voltage -
Standard: 90 to 132 V.
Option 40: 198 to 250 V.
Power Consumption - Less
than 220 VA.
Frequency - 48 to 66 Hz.
CRT - 5.5 in.
Physical Characteristics
|
Dimensions
(approx.)
|
mm
|
in.
|
|
Height
|
177
|
7
|
|
Width
|
330
|
13
|
|
Depth
|
450
|
17.7
|
|
Weight
|
kg
|
lbs.
|
|
R3261C/CN/D
|
15
|
33
|
|
R3361C/CN/D
|
17
|
37.5
|
K475
- Stable/Mobile Platform
- Tiltable Shelf for Optimum Viewing
- Work-bench Height
- Heavy-duty Nylon Instrument Straps
- Surge and Noise Suppression
- Surge-protected Four Outlet Power Strip Accessory
- Shipped Ready to Assemble
|