HP 16533A
250MHz Digital Oscilloscope Card
The HP 16533A digitizing oscilloscope module offers the
advantages of a full-featured deep memory color digitizing oscilloscope
integrated into your logic analyzer. It offers digitizing advantages such
as autoscale, automatic measurements, powerful triggering, negative-time
viewing, voltage markers and time markers.
Features
- Use
as a Standalone Scope with Many Channels
- Capture
up to 32K on each of eight analog channels simultaneously (single
time base).
- Measure
slow and fast events with additional oscilloscope modules to create
a multiple time base digitizing oscilloscope.
- For
large channel count measurements, configure as many as 20 scope
channels in a single system.
- Combine
Scope with Other Logic Analysis Modules
You can arm or trigger the oscilloscope from any module in the logic
analysis system (HP 16500B) to capture and display the analog events
that affect your digital system.
- Autoscale,
Auto-Measure, Voltage and Time Markers and Colors Save You Time
- Select
autoscale and the scope adjusts the time, voltage and trigger levels
instantly for a stable display of your waveforms.
- Use
automatic measurements to analyze a signal's behavior easily.
- Independent
voltage and time markers can be used to measure voltage and timing
relationships.
- Automatic
marker placement and statistics allow you to characterize a circuit
quickly
- Waveforms
are independently colored for fast and easy identification
- Flexible
Display Modes Help You Find Signal Problems
- Capture
random signal variations or metastable states with the accumulate
mode
- Filter
out noise with the average mode
- Show
true single-shot events with the single mode
- Scan
the waveform easily with the connect-the-dots feature
- Analyze
differential waveforms with the A-B mode.
Specifications:
HP 16533A DC to 250 MHz
HP 16534A DC to 500 MHz
DC Offset Accuracy: +/-(1% of offset + 2% of full scale)
DC Voltage Measurement Accuracy: +/-(1.25% of full scale + offset accuracy
+ 0.016 div)
Trigger Sensitivity:
10 mV/div 4 mV/div
DC to 50 MHz 0.25 div 0.63 div
50 MHz to 500 MHz 0.5 div 1.25 div
Input Resistance: 1 MOhm +/- 1%
50 Ohm +/- 1%
Maximum Sampling Rate:
HP 16533A 1 GSa/s
HP 16534A 2 GSa/s
Number of Channels: 2 to 8 using the same time base and trigger.
Up to 10 channels may be installed in a single
HP 16500B frame, or up to 20 in a single system
using an HP 16501A expansion frame.
Waveform Record Length: 32768 points
Vertical (Voltage)
(Characteristics refer to the input at the BNC connector)
Vertical Sensitivity Range: 4 mV/div to 10 V/div in 1:2:4 steps
Vertical Resolution: 8 bits over4 vertical divisions
Rise Time (calculated from bandwidth): 700 ps
Probe Attenuation: Any integer ratio from 1:1 to 1000:1 factor
Channel-to-Channel Isolation (with
channel sensitivities equal):
DC - 50 MHz 40 dB
50 MHz - 500 MHz 30 dB
Maxiumum Safe Input Voltage:
1 MOhm +/-250 V DC + peak AC (< 10 kHz)
50 Ohm 5 Vrms
Horizontal (Time)
Time Base Ranges: 0.5 ns/div to 5 s/div
Time Base Resolution: 10 ps
Delay Range:
Pretrigger 81.8 s - 5 divisions
Posttrigger 2.5E3 seconds
Trigger
Trigger Level Range: Within display window (vertical offset +/-2 divisions)
Trigger Modes:
Immediate Triggers immediately after arming condition is met
Edge Triggers on rising or falling edge of channel 1 or 2
Pattern Triggers on entering ro exiting a specified pattern across
both channels
Auto Condition Self-triggers if trigger is not satisfied within approx.
50 ms after arming
Events Delay The trigger can be set to occur on the nth occurence of a
pattern (n <=32000)
Intermodule Arms another measurement module or activates a trigger
output on the rear panel
BNC connector when the trigger condition is met
|