HP 34905A
Dual 4-Channel RF Multiplexer
HP 34905A Product Details
The
Agilent 34905A RF multiplexer offers broadband switching capabilities
for high-frequency and pulsed signals. Use it to route test signals between
your device under test and your signal generator, oscilloscope, spectrum
analyzer, or other instrumentation.
The RF multiplexers are arranged as two independent 1x4 multiplexers,
each with a common shield and a switched center conductor. Connections
can be made directly to SMB inputs with 2 GHz usable bandwidth, or to
the BNC-to-SMB adapters provided with 1 GHz bandwidth. Multiple banks
may be cascaded together for applications requiring even larger topologies--create
a stubless 16:1 multiplexer in a single frame.
Features:
- Dual
4 Channel RF Multiplexer
- 2
GHz bandwidth, 50 ohm
- BNC
to SMB adapter
cables included
Specifications:
|
Agilent
34905A 50 Ohm
Dual 4-Channel
RF Multiplexer |
| GENERAL
|
|
|
Number of Channels |
Dual 1 x 4, 50 Ohm |
|
Open/Close Speed |
60/s |
| INPUT
|
|
|
Voltage (dc, ac rms) |
42V |
|
Current (dc, ac rms) |
0.7A |
|
Power (W, VA) |
20W |
| DC
CHARACTERISTICS |
|
|
Offset Voltage |
<6µV
|
|
Initial Closed Channel R |
<0.5
Ohm |
|
Isolation ch-ch, ch-earth |
>1
G Ohm |
| AC
CHARACTERISTICS |
|
|
Bandwidth |
2 GHz[1] |
|
Insertion Loss |
-0.1 dB |
|
Insertion Loss |
-0.4 dB |
|
Insertion Loss |
-0.6 dB |
|
Insertion Loss |
-1 dB |
|
Insertion Loss |
-1.2 dB |
|
Insertion Loss |
-3 dB |
|
SWR 10 MHz |
1.02 |
|
SWR 100 MHz |
1.05 |
|
SWR 500 MHz |
1.20 |
|
SWR 1 GHz |
1.20 |
|
SWR 1.5 GHz |
1.30 |
|
SWR 2 GHz |
1.40 |
|
ch-ch Cross Talk |
-100 dB |
|
ch-ch Cross Talk |
-85 dB |
|
ch-ch Cross Talk |
-65 dB |
|
ch-ch Cross Talk |
-55 dB |
|
ch-ch Cross Talk |
-45 dB |
|
ch-ch Cross Talk |
-35 dB |
|
Risetime |
<300
ps |
|
Signal Delay |
<3
ns |
|
Capacitance HI - LO |
<20
pF |
|
Capacitance LO -Earth |
--------- |
|
Volt-Hertz limit |
1010 |
| OTHER
|
|
|
Switch Life No Load (typical) |
5M |
|
Switch Life Rated Load (typical)[2] |
100k |
|
Temperature Operating |
0°C to 55°C |
|
Temperature Storage |
-20°C to 70°C |
|
Humidity (non-condensing) |
40°C / 80% RH |
[1] Bandwidth direct to card SMB connectors