HP 35665A Product Details

The Agilent 35665A is a 2-channel Dynamic Signal Analyzer providing measurement solutions in vibration, acoustics, and control systems. Itsmajor options are computed order tracking, real-time octave, swept sine, curve fit and synthesis, arbitrary source, and Instrument BASIC.

Features:

  • 1-channel at 102.4 kHz, 2 channels at 51.2 kHz

  • 100, 200, 400, 800 lines of resolution

  • Built in 3.5" floppy disk

  • Tachometer input

  • Source: Random, burst random, pink noise, sine,
    swept-sine, arbitrary, periodic chirp, burst chirp

  • 3.2 M Sample time-capture buffer (optional)

  • Measurements: linear spectrum, cross-spectrum,
    power spectral density, frequency response,
    time waveform, auto-correlation, cross-correlation,
    histogram, PDF, CDF

  • Optional real-time octave, computed order tracking, swept-sine, curve fit, synthesis, and arbitrary source.

Specifications:

Frequency
RANGE One Channel Mode (CH. 1): 102.4 kHz Two Channel Mode (Ch 1 & 2): 51.2 kHz RESOLUTION 100, 200, 400 or 800 lines resolution Frequency resolution = frequency span/number of lines resolution Minimum frequency resolution One Channel Mode: 244 uHz Two Channel Mode: 122 uHz SPANS One Channel Mode: 102.4 kHz to 0.19531 Hz Two Channel Mode: 51.2 kHz to 0.097656 Hz ACCURACY ±30 ppm Real Time Bandwidth Fast Average On: >=12.8 kHz MEASUREMENT RATE 401 point FFT display, fast average on One Channel mode: >=33 averages/second (typical) Two channel mode: >=15 averages/second (typical) DISPLAY UPDATE RATE: >=8 updates/second (typical) 401 point FFT display, fast average off
Amplitude/Phase
INPUT RANGES (full scale): +27 dBVrms (31.7 Vpk) to -51 dBVrms (3.99 mVpk) in 2 dB steps AMPLITUDE RESOLUTION: 0.016% of full scale (typical) ABSOLUTE AMPLITUDE ACCURACY: ±2.92% (0.25 dB) of reading ±0.025% of full scale CROSS-CHANNEL GAIN ACCURACY Source = Full scale: ±0.04dB Source = -20dB full scale: ±0.08dB CROSS-CHANNEL PHASE ACCURACY Source = full scale: ±0.5 deg Source = -20 dB full scale: ±0.5 deg
Dynamic Range
FULL SPAN FFT NOISE FLOOR Input Range 27 to -35 dBVrms: <-76 dBfs (-85 dBfs typical) Input Range -37 to -51 dBVrms: <(-112 - Range) dBfs SPURIOUS FREE DYNAMIC RANGE: <-72 dBfs (Includes Spurs, Harmonic Distortion, Intermodulation Distortion, Alias Products) Zoom Mode Alias Responses 1.5% to 98.5% of the Frequency Span: <-72 dBfs Lower and Upper 1.5% of Frequency Span: <-65 dBfs INPUT NOISE LEVEL Above 1280 Hz: <-140 dBVrms / (Sq. R)Hz 160 Hz to 1.28 kHz (6.4kHz span): <-130 dBVrms / (Sq. R)Hz

Input
INPUT RANGES: -51 dBVrms to +27 dBVrms INPUT IMPEDANCE: 1M ohm ±30% < or = 100 pico Farad Low Side to Chassis Impedance Floating Mode: 1M ohm ±30% (typical) <0.01 micro Farad (typical) AC COUPLING: <3 dB Rolloff at 1Hz COMMON MODE REJECTION -51 dBvVrms to -11 dBvVrms Ranges: >80 dB (typical) -9 dBVvrms to +9 dBVrms Ranges: >60 dB (typical) +11 dBvVrms to +27 dBVrms Ranges >40 dB (typical) COMMON MODE RANGE (floating mode): ±4 V pk IEPE SIGNAL CONDITIONING Current Source: 4.25 ±1.5 mA Open Circuit voltage: +26 to +32 Vdc
Trigger/Tachometer
TRIGGER MODES: Internal, External TTL, Source, HP-IB MAXIMUM TRIGGER DELAY Post Trigger: 8191 seconds Pre Trigger: 8191 samples PULSES PER REVOLUTION: 0.5, 1 to 2048 (Integer Steps) RPM ACCURACY: ±100 ppm TRIGGER LEVEL RANGE Low Range: -4V to +4V High Range: -20V to +20V TRIGGER LEVEL RESOLUTION Low range: 100 mV High Range: 500 mV RPM RANGE: 5 < RPM < 220,759 MAXIUM TRIGGER PULSE RATE: 800 kHz
Source Output
AMPLITUDE RANGE: ±5 V peak AMPLITUDE RESOLUTION Voltage >/= 0.2 Vrms: 2.5 mVpeak Voltage <0.2 Vrms: 0.25 mVpeak RESIDUAL DC OFFSET: ±10.0 mV OUTPUT IMPEDANCE: <5 ohms SOURCE TYPE: Sine, random, pink noise, chirp Options
COMPUTED ORDER TRACKING Pulses per Revolution: 0.5, 1 to 2048 (integer steps) (Max Order x Max RPM) / 60 On line (real-time) 1 Channel Mode: <= to 25,600 Hz 2 Channel Mode: <= to 12,800 Hz Capture playback 1 Channel Mode: <= to 102,400 Hz 2 Channel Mode: <= to 51,200 Hz Specified for 5 < RPM < 220,759 RPM (on line) 5 < RPM < 491,519 RPM (capture playback); and Number of Orders < or = 200. Software Limits at 5 < or = RPM < or = 491,519; and Number of Orders < = 200 Delta Order: 1/128 to 1/1 Resolution (Maximum Order) / (Delta Order): <200 RPM Accuracy: ±100 ppm Maximum RPM Ramp Rate: 750 RPM/second (typical) 1 and 2 Channel Mode 1000 to 10,000 RPM Run-up Maximum Order = 10 Delta Order = 0.1 RPM Step = 30 REAL TIME OCTAVE ANALYSIS - OPTION 1D1 STANDARDS ANSI Standard S1.11 - 1996, Order 3, Type 1-D, Extended and Optional Frequency Ranges IEC Standard 651-1979 Type 0 Impulse, and ANSI S1.4 1/1 OCTAVE RANGE (at centers): 0.0613 Hz to 16 kHz 1/3 OCTAVE RANGE (at centers): 0.08 Hz to 32 kHz 1/12 OCTAVE RANGE (at centers): 0.09145 Hz to 12.338 kHz 1/3 OCTAVE UPDATE RATE 1 Channel Mode 16 kHz Bandwidth 12 Updates/Second (Typical) 32 kHz Bandwidth 5 Updates/Second (Typical) 2 Channel Mode 8 Updates/Second (Typical) (16 kHz, 32 kHz Bandwidths) 1/3 OCTAVE DYNAMIC RANGE 2 Second Stable Average: >80dB as per ANSI S1.11-1986 Totalpower limited by the for full scale ranges from input noise level 50 m Vpeak to 31.7 Vpeak STABLE AVERAGE RANGE One Channel Mode Real-Time Bandwidth = 32 kHz Two Channel Mode Real-Time Bandwidth = 16 kHz Ordering Information: SWEPT SINE MEASUREMENTS - OPTION 1D2 DYNAMIC RANGE: >130 dB (typical) Input auto range on 100 msec integration SWEEP TIMES: 51.2 Hz to 51.2 kHz span 11 dBV rms source level 100 ms integration time approximately 76 sec (typical) 10 ms integration time approximately 25 sec (typical) 5 cycle integration time approximately 16 sec (typical) 100 cycle integration time approximately 25 sec (typical) CROSS-CHANNEL AMPLITUDE/PHASE ACCURACY Ch. 1 range = Ch. 2 range: ±0.04 dB ±0.5 deg

Agilent HP 35665A 2-Channel DC to 102.4 kHz

      Opt 35665A 2-Channel DC to 102.4 kHz Dual Channel Dynamic Signal Analyzer
      Opt 0BU Additional instrument BASIC manuals
      Opt 1C2 Add HP instrument BASIC
      Opt 1CM Rack Flange Kit
      Opt 1D0 Computed order tracking
      Opt 1D1 Real time octave measurements
      Opt 1D2 Swept sine measurement
      Opt 1D3 Curve fit/synthesis
      Opt 1D4 Arbitrary source
      Opt 1F0 PC-style keyboard - U.S. version
      Opt ANA Increase RAM to 8 MB total


 
 