HP 54616C
500MHz Digital Oscilloscope
The HP 54616C color oscilloscope continues in the tradition of Hewlett-Packard's
popular HP 54600-series of digitizing scopes by delivering the comfortable
feel of an analog oscilloscope with the power of a digital architecture
-- and now, with color. In addition, this oscilloscope offers an incredibly
high level of digitizing performance to give you confidence in your critical
measurements at a fraction of the price that you might expect.
The Feel of Analog: When you're troubleshooting, you want to stay
focused on two things: your circuit and the scope's display. You don't
want to waste time pressing buttons or waiting for the scope to update.
That's why the HP 54616C has front-panel controls that look and work like
the controls on your old analog scope. Color provides the ability to identify
and interpret the information on the display quickly, without taking attention
off the task at hand.
The Power of Digital: Digitized waveforms open up entirely new
possibilities, such as pretriggering, autoscale to automatically set voltage,
time and trigger parameters, the ease and accuracy of automatic measurements,
saving and recall of traces, hardcopy output to printers, and remote control
with your computer.
Features:
- Color display
- 2 GSample/sec simultaneously on 2 channels
- 500 MHz analog bandwidth
- 500 MHz single shot bandwidth
- 1 ns peak detect
- Optional expansion modules for FFT, waveform storage, and remote control
- Sweep speeds from 1 ns/div to 5 s/div
Specifications:
Primary Vertical Specifications
Channels 2
Bandwidth (-3dB) dc to 500 MHz
ac coupled 10 Hz to 500 MHz
Sensitivity 2 mV/div to 5 V/div
Accuracy ±2%
Vernier Accuracy (2) ±2%
Rise Time <700 ps (calculated)
Input Coupling dc, ac, & ground
Input Impedance 1 Mohm, approx. 9 pF or 50 ohm
selectable
Secondary Vertical Specifications
Bandwidth Limit ≈ 30 MHz
Inversion CH 1 & CH 2
CMRR ≈ 20 dB at 50 MHz
Dynamic Range ±12 div from center screen
Maximum Input 250 V (dc + peak ac) or
5 Vrms in 50 ohm mode
50-ohm Protection Protects 50 ohm load from
excessive voltage
Time Skew Adjustable over a range of ±25 ns to
remove effects of cabling
Probe Sense Automatic readout of
1X, 10X, 20X, and 100X probes
Voltage Measurement Accuracy Specifications
Single Cursor Vertical Accuracy ±1.2% of full
scale ± 0.5% of position value
Dual Cursor Vertical Accuracy ±0.4% of full scale
Math Functions CH 1 + or - CH 2
Horizontal System Specifications
Main & Delayed
Main Sweep Range 5 s/div to 1 ns/div
Delayed Sweep Range Up to 200X main sweep, as
fast as 1 ns/div
Accuracy ±0.005%
Resolution 20 ps
Delay Jitter 1 ppm
Pretrigger Delay = 10 div
(Negative time)
Posttrigger Delay At least 2,560 div or 50 ms,
(Trigger to start of sweep) not to exceed 100 s
Roll Mode At sweep speeds of 500 ms/div and
slower, waveform data moves across
the display from right to left with
no dead time.
Display can be free running (nontrig-
gered) or triggered to stop on a
trigger event.
Time Measurement Accuracy
Cursor accuracy ± 0.01%
(delta-t & 1/delta-t) ± 0.2% of full scale, ± 200 ps
Internal Triggering Specifications
Sensitivity (Ch 1 & 2)
dc to 25 MHz: 0.5 div or 3.5 mV
dc to max. bandwidth: 1 div or 7 mV
Sources Channels 1 & 2, line and external
Coupling ac, dc, HF reject, LF reject, & noise
reject (LF & HF: -3db at
approx. 50 kHz)
Modes Auto, Autolevel, Normal, Single, & TV
External Triggering Specifications
Range ± 18 V
Sensitivity dc to 100 MHz: < 75 mV
dc to 500 MHz: < 150 mV
Coupling dc, ground
Input R&C 1 Mohm, ~12 pf or 50 ohm selectable
Maximum Input 250 V (dc + peak ac) or
5 Vrms in 50-ohm mode
TV Triggering Specifications
TV line and field. 0.5 div of composite sync for stable display
(Ch 1 & Ch 2)
Line Counting Delay time calibrated in
NTSC and PAL line numbers
All Field Trigger Oscilloscope triggers on the vertical
sync pulse in both fields, allowing
use with noninterlaced video.
Trigger Holdoff
Adjustable from 200 ns to approximately 13 s
X-Y Operation
Bandwidth X & Y same as vertical system
Phase Difference ± 3° at 10 MHz
Display System
Display 5.8 inch Active matrix color LCD
Resolution 255 vertical by 500 horizontal points
Controls Front-panel intensity control
Graticule 8 X 10 grid or frame
Autostore Saves previous sweeps in half bright
display and the most recent sweep in
full bright.
Acquisition System Specifications
Max Sample Rate 2 GSa/s
Resolution 8 bits
Simultaneous Channels 2
Record Length 5,000
Usable Single-Shot Bandwidth 500 MHz
Peak Detect 1-ns glitch capture
Average Number of averages selectable
at 8, 64, 256
Advanced Functions
Automatic Measurements Measurements are continously updated
Voltage Vavg, Vrms, Vpp, Vtop,
Vbase, Vmin, & Vmax
Time Frequency, Period, +Width, -Width,
Duty Cycle, Rise Time, & Fall Time
Cursors Manually or automatically placed
Setup Functions
Autoscale Sets the vertical and horizontal
deflection and the trigger level.
Save/Recall 10 front-panel setups
Trace Memory Two volatile pixel memories
Environmental Specifications
Line Voltage Range 100 Vac to 240 Vac
Line Voltage Selection Automatic
Line Frequency 45 Hz to 440 Hz
Max Power Consumption 220 VA
Temperature Operating: 0 °C to +55 °C
Nonoperating: -40 °C to +70 °C
Humidity Operating: 95% RH at 40°C for 24 Hrs
Nonoperating: 90% RH at 65°C for 24 Hrs
Altitude Operating: to 3,048 m (10,000 ft)
Nonoperating: to 12,192 m (40,000 ft)
Size (excluding handle) Height: 172 mm (6.8 in)
Width: 322 mm (12.7 in)
Depth: 317 mm (12.5 in)
Weight 6.2 kg (14 lbs)
Safety CSA Certification, IEC 348
|