HP 83236A
PCS Adapter
HP 83236A Product Details
The Agilent 83236A PCS Interface was designed to extend Agilent 8921A/D
TDMA and CDMA measurement capability to the 1850 - 1990 MHz U.S. PCS band.
The Agilent 83236A hardware can be integrated with existing RF test
systems in the field. Adding the Agilent 83236A to existing Agilent 8921A/D
systems is a low cost way to obtain PCS - band measurement capability.
Features:
- Cost-effective PCS upgrade to Agilent 8921A/D Cell Site Test Sets.
- Internal power meter to maintain power measurement accuracy and speed
at PCS-band frequencies.
Specifications:
Generator Output Path (RF In/Out and RF Out Only connector)
Note: To meet generator output path specifications, the input signal
must be from an Agilent 8921A/D RF test set with the following characteristics;
Input Frequency Range: 824 MHz to 894 MHz and Input Level Range: –70 dBm
to –7 dBm.
Frequency Range: Through Path: 824 to 849 MHz 869 to 894 MHz
Conversion Path: 1850 to 1990 MHz Typical Frequency Settling
Time: <10 msec Output Level Range RF In/Out Connector: –130
to –20 dBm RF Out Only Connector:–130 to –10 dBm Output Level
Settling Time: <100 msec Level Accuracy: +/- 1.8 dB, at 25
degrees C +/- 10 degrees C +/- 2.0 dB, from 0 to 55 degrees C Typically
+/- 1.0 dB
Analyzer Input Path
Frequency Range: Through Path: 824 to 849 MHz 869 to 894 MHz
Conversion Path: 1850 to 1990 MHz Maximum Input Level:
40 dBm (10 W) Integrated Spurious and Phase Noise: <–57 dBc in
a 100 Hz to 32 kHz bandwidth Spurious Level: <–60 dBc at ±5 kHz
from carrier
Power Measurement Frequency Range: 824 to 849 MHz 869 to 894
MHz 1850 to 1990 MHz Measurement Range: –13 dBm to 40 dBm (50
mW to 10 W) Accuracy: ±5% ±2.5 microWatts (at 23 ±10 degrees
C after power meter zero and calibration) ±10% ±2.5 microWatts from
0 to 50 degrees C Resolution: 0.01 dB or 10 mW
|