Agilent HP 85110A Pulsed RF S-Parameter Test Set, 2 GHz to 20 GHz
The Agilent 85110A test set is specially configured for operation
in pulsed-RF measurement systems. Four 90 dB step attenuators protect
each input of the fundamentally-mixed down converter to allow measurement
of test devices with output power of 20 W CW. Special test set options
are available from .5 to 20 GHz and 2 to 50 GHz.
Features:
- 2 to 20 GHz frequency range
- For use in pulsed-RF measurement systems
- Measurement of devices with 20 W CW output power
Specifications:
Frequency Range: 2 to 20 GHz
Test Ports (port 1 or 2)
Nominal Operating Power Level:
0 to -3 dBm
Connector Type: 3.5 mm, male
Impedance, DC Bias: 50 ohm nominal, 500 mA, 40 Vdc maximum
Attenuation Range (Incident Signal):
0 to 90 dB, in 10 dB steps
RF Input Connector (Rear Panel):
Maximum Input Power: +14 dBm
Connector Type: 3.5 mm, female
Dynamic Range1 (for transmission measurements,
independent of duty cycle)
Frequency Range (GHz)
2 to 8 8 to 18 18 to 20
Maximum Power (dBm)
Measured at Port 22: +11 +11 +11
Reference Power (dBm)
at Port 1 (nominal): +0 -1 -2
Minimum Power (dBm)
Measured at Port 2
Pulsed: -64 -63 -62
CW: -78 -78 -77
Receiver
Dynamic Range (dB)
Pulsed: 75 74 73
CW: 89 89 88
System
Dynamic Range (dB)
Pulsed: 64 62 60
CW: 78 77 75
1Limited by compression level and system
noise floor. Noise floor is measured with full two-port error correction,
1024 averages.
2This maximum power measurement assumes that the Agilent
85110A test set has its internal step attenuators set to 0 dB. The
test can handle up to 20 W (+43 dBm) or power if the step attenuators
are activated and an isolator is installed (in the port 2 real panel
link).
Measurement Port Characteristics
Frequency Range (GHz)
2 to 8 8 to 18 18 to 20
Residual (dB)
Directivity: 44 44 44
Source Match: 33 31 31
Load Match: 44 44 44
Reflection Tracking: ±0.003 ±0.006 ±0.006
Transmission Tracking: ±0.044 ±0.084 ±0.094
Crosstalk: 64 62 61
Pulsed-RF Detectors
Equivalent Measurement Bandwidth (-3 dB): 1.5 MHz
Aperature Uncertainty: <1 nanosecond, typical
Transition Time (10% to 90%): 300 nanoseconds
Trigger Level: TTL (falling edge), external
Trigger Width: 100 nanoseconds, minimum
Minimum Time Display: 5 microseconds
Maximum Time Display: 40 milliseconds
Ordering Information:
Agilent HP 85110A Pulsed RF S-Parameter Test Set,
2 GHz to 20 GHz
| Agilent 85110A |
Pulsed RF S-Parameter Test Set, 2 to 20 GHz |
| Opt 001 |
Add IF switching circuitry |
| Opt 908 |
Rack flange kit |
| Opt 910 |
Additional operating and service manual |
| Opt 913 |
Rack flange and handle kit |
|