Home   |   Contact Us   |   Shopping Cart   |   Quote Cart  0 items   |   Login
MetricTest Equipment Logo   Advanced
1.800.417.4370
 

Agilent HP 8514B S-Parameter Test Set, 0.045GHz to 20GHz, 3.5 mm

Several S-parameter tests sets are available for the Agilent 8510C network analyzer for broadband coaxial measurements from 45 MHz to 50 GHz. The Agilent 8514B test set has an architecture that develops a separate reference channel for each incident port. RF switching is done with a built-in electronic switch. For active device measurements, the test set includes the ability to apply dc bias (external) to the test port center conductors. Also available are two 90 dB step attenuators, which allow control of the port 1 and port 2 signal levels.

Features:

  • 45 MHz to 20 GHz
  • Separate reference channel for each incident port
  • Built-in electronic switch
  • 3.5 mm connectors

Specifications:

Frequency Range: 0.045 to 20 GHz
Test Ports (port 1 or 2)
Nominal Operating Power Level:
2 to -6 dBm
Connector Type: 3.5 mm, male
Impedance, DC Bias: 50 ohm nominal, 500 mA, 40 Vdc maximum
Attenuation Range (incident signal)
0 to 90 dB, in 10 dB steps

RF Input Connector (rear panel)
Maximum Input Power: +16 dBm
Connector Type: 3.5 mm, female

Dynamic Range (for transmission measurements)
Frequency Range (GHz)
.045 to 2 2 to 8 8 to 20
Maximum Power (dBm)
Measured at Port 2: +20 +11 +10

Reference Power (dBm)
at Port 1 (nominal): +2 -2 -6

Minimum Power (dBm)
Measured at Port 2: -66 -95 -95
Receiver
Dynamic Range (dB): 86 106 105
System
Dynamic Range (dB): 68 93 89
Measurement Port Characteristics
Frequency Range (GHz)
.045 to 2 2 to 8 8 to 20
Residual (dB)

Directivity: 48 44 44
Source Match: 40 33 31
Load Match: 48 44 44
Reflection Tracking: ±0.003 ±0.003 ±0.006
Transmission Tracking: ±0.017 ±0.044 ±0.084
Crosstalk: 89 115 110
Raw (typical) (dB) (before calibration/error correction)

Directivity: 23 23 14
Source Match: 17 15 11
Load Match: 17 15 11

Ordering Information:

Agilent HP 8514B S-Parameter Test Set, 0.045GHz to 20GHz, 3.5 mm

Agilent 8514B S-Parameter Test Set, 0.045 GHz to 20 GHz, 3.5 mm
      Opt 001 Add IF switching
      Opt 002 Delete step attenuators and bias tees
      Opt 003 High forward dynamic range
      Opt 1BN Mil std 45662A calibration certification
      Opt 1BP Mil std 45662A calibration w/test data
      Opt 908 Rack flange kit, front handles detached
      Opt 910 Extra operating and service manual
      Opt 913 Rack flange kit, front handles attached



 
 
  Company    |   Contact Us   |   Site Map 1.800.417.4370
 
 
 
 