Agilent HP 8515A S-Parameter Test Set, 45 MHz to 26.5 GHz, 3.5 mm
Several S-parameter tests sets are available for the Agilent 8510C
network analyzer for broadband coaxial measurements from 45 MHz
to 50 GHz. The Agilent 8515A test set has an architecture that develops
a separate reference channel for each incident port. RF switching
is done with a built-in electronic switch. For active device measurements,
the test set includes the ability to apply dc bias (external) to
the test port center conductors. Also available are two 90 dB step
attenuators, which allow control of the port 1 and port 2 signal
levels.
Features:
- 45 MHz to 26.5 GHz
- Separate reference channel for each incident port
- Built-in electronic switch
Specifications:
Frequency Range: 0.045 to 26.5 GHz
Test Ports (port 1 or 2)
Nominal Operating Power Level:
-5 to -25 dBm
Connector Type: 3.5 mm, male
Impedance, DC Bias: 50 ohm nominal, 500 mA, 40 Vdc
maximum
Attenuation Range (incident signal)
0 to 90 dB, in 10 dB steps
RF Input Connector (rear panel)
Max Input Power: +14 dBm
Connector Type: 3.5 mm, female
Dynamic Range (for transmission measurements)
Frequency Range GHz
.045 to 2 2 to 8 8 to 20 20 to 26.5
Maximum Power (dBm)
Measured at Port 2: +2 +3 +3 -1
Reference Power (dBm)
at Port 1 (nominal): -5 -9 -14 -25
Minimum Power (dBm)
Measured at Port 2: -101 -102 -101 -99
Receiver
Dynamic Range (dB): 103 105 104 100
System
Dynamic Range (dB): 96 93 87 74
Measurement Port Characteristics
Frequency Range (GHz)
.045 to 2 2 to 8 8 to 20 20 to 26.5
Residual (dB)
Directivity: 48 50 50 50
Source match: 40 50 50 50
Load match: 48 50 50 50
Reflection Tracking: ±0.003 ±0 ±0 ±0
Transmission Tracking: ±0.009 ±0.004 ±0.009 ±0.01
Crosstalk: 117 114 108 95
Raw (typical) (dB) (before calibration/error correction)
Directivity: 24 24 28 27
Source Match: 23 23 16 14
Load Match: 23 23 16 14
Ordering Information:
Agilent HP 8515A S-Parameter Test Set, 45 MHz to
26.5 GHz, 3.5 mm
| Agilent 8515A |
S-parameter Test Set, 45 MHz to 26.5 GHz, 3.5 mm |
| Opt 001 |
Add IF switching circuitry |
| Opt 002 |
Delete step attenuators & bias networks |
| Opt 908 |
Rack flange kit |
| Opt 910 |
Additional operating and service manual |
| Opt 913 |
Rack flange and handle kit |
| Opt W03 |
Convert to standard 3 mo. on-site warr. |
| Agilent 8515A |
Mil std 45662A calibration |
| Opt 1BN |
MIL-STD 45662A Calibration |
| Opt 1BP |
MIL-STD 45662A Calibration with Test Data |
|