Agilent HP 86120B Multi-Wavelength Meter
The Agilent 86120B Multi-Wavelength Meter is a Michelson interferometer-based
instrument that measures wavelength and optical power of laser light
in the 700 to 1650 nm wavelength range. Simultaneous measurements
of multiple laser lines are performed allowing measurements of WDM
(wavelength division multiplexed) signals and multiple lines of
Fabry-Perot lasers.
Features:
- Multiple and average wavelengths measured
- Peak and total powers measured
- Automatic signal-to-noise ratios calculated
- Power and wavelength of up to 100 channels
- Built-in routines for channel spacing and drift
- Accurate enough for use in the lab and rugged enough for use
in the field
Specifications:
Wavelength
Range
700 to 1650 nm (182 to 428 THz)
Absolute accuracy, laser lines separated by >=30 GHz ±3
ppm
(±0.005 nm at 1550 nm, ±0.004 nm at 1310 nm)
Differential accuracy(characteristic) ±2 ppm
Minimum resolvable separation (equal power lines input)
20 GHz (0.16 nm at 1550 nm, 0.11 nm at 1300 nm)
Display resolution
0.001 nm, normal update mode
0.01 nm, fast update mode
Power
Calibration accuracy
±0.5 dB (at ±30 nm from 780, 1310, and 1550 nm)
Flatness, 30 nm from any wavelength
1200 to 1600 nm ±0.2 dB
700 to 1650 nm ±0.5 dB
Linearity ±0.3 dB
Polarization dependence,
1200 to 1600 nm ±0.5 dB
700 to 1650 nm ±1.0 dB
Display resolution 0.01 dB
Sensitivity
700 to 800 nm, singleline input -20 dBm
800 to 1200 nm, single line input -25 dBm
1200 to 1600 nm, single line input -40 dBm
1600 to 1650 nm, single line input -30 dBm
700 to 1650 nm, multiple lines input
30 dB below total input power, but not
less than single line input sensitivity
Selectivity
2 lines input separated by >=100 GHz 25 dB
2 lines input separated by >=30 GHz 10 dB
Input Power
Maximum displayed level (sum of all lines input)+10 dBm
Maximum safe input level (sum of all lines input)+18 dBm
Weight: 9 kg (19 lb)
Dimensions: (H x W x D) 140 x 340 x 465 mm; 5.5 x 13.4 x
18.3 in
Ordering Information:
Agilent HP 86120B Multi-Wavelength Meter
| Agilent 86120B |
Multi-Wavelength Meter |
| Opt 011 |
Replace FC/PC connector interface with Diamond HMS-10/HP
connector interface |
| Opt 013 |
Replace FC/PC connector interface with DIN 47256 connector
interface |
| Opt 014 |
Replace FC/PC connector interface with ST connector interface
|
| Opt 017 |
Replace FC/PC connector interface with SC connector interface
|
| Opt 022 |
Replace flat physical contact interface with angled physical
contact interface |
| Opt 412 |
Adds an external 10 dB attenuator with flat FC/PC connector
|
| Opt 417 |
Adds an external 10 dB attenuator with angled FC/PC connector
(requires option 022) |
| Opt AXE |
Rack flange kit with handles |
| Opt IA4 |
Rackmount kit without handles |
| Opt UK5 |
Nylon carrying case with shoulder strap |
| Opt UK6 |
Commercial calibration certificate with test data |
| Opt UK7 |
Hard carrying case |
|