Agilent HP 87511B S-Parameter Test Set

The 87511A/B S-Parameter Test Sets provide the capability to measure S-parameters of 2-port devices from 100kHz to 500MHz with an Agilent network analyzer.

Features:

  • Frequency is 100kHz to 500MHz
  • Test Port Impedance is 50 Ohm for 87511A and 75 Ohm for 87511B

Specifications:

Parameter

Agilent 87511A

Agilent 87511B

Type

Frequency
Range

100kHz to 500MHz

Specification

Directivity
Port 1 &2

> 30dB 100kHz to 300KHz
> 40dB 300kHz to 200MHz
>35dB 200MHz to 500MHz

> 30dB 100kHz to 300KHz
> 33dB 300kHz to 500MHz

Port match:
R,A,B, RF IN

> 20dB

Typical

Switch
Repeatability

+/- 0.03dB

Specifications

Source Match

> 15dB, 100kHz to 300kHz
> 20dB, 300kHz to 500MHz

Frequency
Response

+/- 1dB

Typical

Insertion Loss
RF IN to Port 1&2
RF IN to R,A,B

13dB
19dB

19dB
31dB

Nominal

Isolation between
Port 1 and 2

110dB

100dB

Typical

Impedance
Port 1&2
RF IN, R,A,B

50 Ohm
50 Ohm

75 Ohm
50 Ohm

Nominal

Connectors:
Port 1 & 2

7mm (50 Ohm)
50 Ohm type-N(f) (option 001)

75 Ohm Type-N(f)

Connectors:
RF IN, R, A, B

50 Ohm type-N(f)

Max. Input Level

+20dBm

Dimensions

432mm x 90mm x 495mm

Weight

5.7kg

Ordering Information:

Agilent HP 87511B S-Parameter Test Set

Agilent 87511A S-Parameter Test Set
      Opt 001 50 Ohm Type-N Test Port
      Opt 1BQ Calibrate to mil standard 45662
      Opt 907 Front Handle Kit
      Opt 908 Rack Mount Kit
      Opt 909 Rack Mount Flange and Handle Kit
      Opt 910 Extra Operation Manual
      Opt UK6 Commercial cal. certificate w/ test data

 

 

 



 
 
  Company    |   Contact Us   |   Site Map 1.800.417.4370
 
 
 
 