Agilent HP 8901A Modulation Analyzer, 150 kHz to 1300 MHz
The Agilent Technologies 8901A modulation analyzer combines the capabilities
of several RF instruments to give complete, accurate characterization
of modulated signals in the 150 kHz to 1300 MHz frequency range. It very
accurately measures modulation and recovers the modulation signal, determines
RF frequency and measures RF power.
Features:
- Measures AM and FM to 1% accuracy
- Measures RF frequency
- Measures RF power
- Low internal noise
- Completely automatic
Specifications:
RF Input
Frequency Range: 150 kHz to 1300 MHz Operating
Level: 12 mV rms to 7 V rms Input Impedance: 50 ohms nominal
Tuning: Manual frequency entry, automatic, or track Acquisition
Time: (automatic operation):1.5s
Frequency Modulation
Rates: 20 Hz to 200 kHz Deviations: To
400 kHz Accuarcy: +/- 2% of reading +/- 1 digit, 20 Hz to 10 kHz
rates, 250 kHz to 10 MHz; +/- 1% of reading +/- 1 digit, 50 Hz to 100
kHz rates, 10 to 1300 MHz
Phase Modulation
Carrier Frequency: 10 to 1300 MHz Rates:
200 Hz to 20 kHz; typically usable from 20 Hz to 100 kHz with degraded
performance Deviation: To 400 radians Accuracy: +/- 3% of
reading +/- 1 digit
Amplitude Modulation
Rates: 20 Hz to 100 kHz Depth: To 99% Accuracy:
+/- 2% of reading +/- 1 digit, 50 Hz to 10 kHz rates, 150 kHz to 10 MHz
+/- 1% of reading +/- 1 digit, 50 Hz to 50 kHz rates, 10 to 1300 MHz
Frequency Counter
Range: 150 kHz to 1300 MHz Accuracy: +/-3
counts of least significant digit +/- reference accuracy
RF Level (Peak Voltage Responding, RMS Sine Wave Power Calibrated)
Range: 1 mW to 1 W Instrumentation Accuracy:
+/- 1.5 dB
Power Reference
Power Output: 1.00 mW, factory set to +/- 0.7%,
traceable to the U.S. National Institute of Standards and Technology Accuracy:
+/- 1.2% worst case (+/-0.9% rss) for one year (0 degree to 55 degrees
C)
Ordering Information:
Agilent HP 8901A Modulation Analyzer, 150 kHz to 1300
MHz
| Agilent 8901A |
Modulation Analyzer, 150 kHz to 1300 MHz |
| Opt 001 |
Rear panel connections instead of front |
| Opt 002 |
High stability timebase, internal |
| Opt 003 |
Rear panel external LO connectors |
| Opt 004 |
48-440 Hz line power |
| Opt 010 |
AM/FM calibrator; accuracy: +/- 0.1% |
| Opt 0Q8 |
Factory delivered service training |
| Opt 907 |
Front handle kit |
| Opt 908 |
Rack flange kit |
| Opt 909 |
Front handle and rack flange kit |
| Opt 910 |
Additional manuals |
| Opt 915 |
Service manual |
| Opt UK6 |
Commercial cal. certificate w/ test data |
|