Agilent HP E1440A VXI21 MHz Synthesized Function/Sweep Generator
The Agilent E1440A function/sweep generator is a C-size, 2-slot,
message-based VXI module. It provides low-distortion sinewaves
and a variety of waveforms for applications requiring high frequency
stability and resolution (11 digits). A programmable relative phase
output synchronized to a companion module is also available. Two
or more E1440A modules can be used for generating multi-phase related
signals.
With this module, you can use the modulation source as an arbitrary
function generator via GPIB to provide user-defined waveforms, or
use the save-recall memory that includes nonvolative memory locations
for simple and rapid access to frequently used test setups. Additionally,
you can produce five different waveforms including: sine, square,
triangle, negative, and positive ramps.
Features:
- Five different waveforms
- Multi-interval sweep and multimarker mode
- Amplitude and phase modulation
- 1 µHz - 60 MHz TTL clock
- High-voltage output option, Isolated (floating) outputs
- Two or more HP E1440A modules for generating phase-related signals
Specifications:
Waveforms
sine, square, triangle, negative and positive Ramps,
dc, TTL clock
Frequency Range
Sine: 1 µHz - 21 MHz Square: 1 µHz
- 11 MHz Triangle: 1 µHz - 11 kHz Ramps: 1 µHz - 11
kHz TTL clock: 1 µHz - 60 MHz Resolution: 11 digits
Accuracy: ±5 ppm of selected value, 20 to 30°C, at time of
calibration with standard frequency reference Stability:
±5 ppm/year, 20 to 30 °C Modulation: AM, PM
Main Signal Output (Typical)
Impedance: 50 Ohm ±1 Ohm Maximum voltage:
42 V pk (AC+DC) (chassis ground to circuit ground, 0 - 10 kHz) Maximum
voltage: ±10 V (floating ground to signal output) Connector:
BNC
Amplitude
(50 ohm all waveforms except TTL clock, without
DC offset) Range: 1 mV to 10 Vp-p in 8 amplitude ranges,
1-3-10 sequence, amplitude can also be set up in Vrms and dBm Resolution:
4 digits (0.03% of full range) Amplitude Accuracy: ±2% FS
Sinewave Spectral Purity
Phase noise: -55 dB for a 30 kHz band centered
on a 20 MHz carrier (excluding ±1 Hz about the carrier) Spurious:
-60 dBc or -85 dBm, whichever is greater Sinewave harmonic distortion:
Frequency Range Harmonic Level 0.1 Hz - 199 kHz -60 dBc 200
kHz - 1.99 MHz -40 dBc 2 MHz - 14.9 MHz -30 dBc 15 MHz - 20 MHz
-25 dBc
Squarewave Characteristics (Typical)
Rise/fall time: <=20 ns (10% to 90%
of p-p output voltage) Overshoot: 5% of p-p amplitude
at full output Triangle/ramp linearity: ±0.05% of full p-p
output voltage for each range (10% - 90%, 10 kHz)
DC Offset Range (50 ohm)
DC only: 0 to ±5 V DC + AC: max.
±4.5 V Resolution: 4 digits
Phase Offset
(related to other Agilent E1440A or equivalent)
Range: ±719.9° Resolution: 0.1° Increment accuracy:
±0.5° Stability: ±1° of phase / °C
Sinewave Amplitude Modulation (Typical)
Modulation depth: 0 - 98% Modulation
frequency range: DC to 350 kHz (1 µHz - 21 MHz carrier frequency)
Envelope distortion: -30 dB for modulation to 80% at 1 kHz
(0 Vdc Offset) Sensitivity: ±5 V peak for maximum modulation
Phase Modulation (Typical)
Sinewave range: ±900°, ±5 V input Sinewave
linearity: ±0.5%, best fit straight line up to ±720° of modulation
range Squarewave range: ±450° Triangle range: ±45°
Positive and negative ramps range: ±90° Modulation frequency
range: DC to 5 kHz
Frequency Sweep
Sweep sequence modes: single, continuous
Sweep function modes: Multi-Interval: (up to 50 different
intervals can be sequenced and repeated in any order in a sequence
that can contain up to 100 intervals) Linear or logarithmic sweep:
(can be set for each interval) Sweep time: Linear:
0.01 s to 105 s Logarithmic: 0.1 s to 105 s Minimum sweep
width: Linear: 0 Hz Logarithmic: 1 decade Maximum
sweep width: full frequency range Minimum sweep rate:
Linear: 0.2 Hz/s Phase continuity: sweep is phase
continuous over the full frequency range of the main output for
all sweep modes Multimarker: Linear sweep only: up
to nine markers can be set in this one dedicated interval Sweep
time: 0.01 s to 105 s Sweep width: from 0 Hz to full
frequency range
Auxiliary Outputs(Typical)
SYNC-OUT TTL: Signal: Phase synchronous
squarewave with same frequency as the main signal output, or 1 µHz
to 60 MHz TTL clock (main signal output switched off) Output
impedance: 50 ohm Connector: BNC and TTL trigger bus
X-Drive 0 to 10 V: Signal: 0 - 100 s sweeps only (proportional
ramp to the entire sweep time) Output impedance: 650 ohm
Output level: 0 to + 10 V (into open circuit) Connector:
BNC Pen lift: Signal: TTL-compatible voltage levels
capable of sinking current from a positive source. Current 200 mA,
voltage 45 V Connector: BNC Marker TTL: Signal:
High-to-low transitions at selected marker frequencies. TTL- and
CMOS-compatible output levels Pulsewidth in multimarker mode:
1 ms Connector: BNC and TTL trigger bus Fan out: 4
REF out 10 MHz Signal: 10 MHz squarewave for phase-locking
additional instruments to the Agilent E1440A. Output impedance:
50 ohm Output levels (into 50 ohm): high level >2 V, low
level <0.2 V AC-coupled output levels: 10 dBm Connector:
BNC
Auxiliary Inputs (Typical)
External REF in 1/10 MHz: (For phase-locking
the Agilent E1440A to an external frequency reference) Signal:
from 0 dBm to 20 dBm into 50 ohm (reference signal must be a
subharmonic of 10 MHz from 1 MHz to 10 MHz) Connector:
BNC or VXI-system clock AM: Input impedance: 10 kOhm
Connector: BNC Max. external voltage: ±15 V PM:
Input impedance: >40 kOhm Connector: BNC Max.
external voltage: ±15 V
Option 001 High-Voltage Output
Frequency range: 1 µHz to 1 MHz Amplitude:
Range: 4 mV to 40 V p-p in eight ranges, 4-12-40 sequence
into 500 ohm, <500 pF load; ranges are four times the standard
instrument ranges, without DC offset Accuracy: ±2% of full
output for each range at 2 kHz Flatness: ±10% relative to
programmed amplitude Sinewave harmonic distortion: Frequency
Range Harmonic Level 10 Hz - 199 KHz - 60 dBc 200 kHz - 1 MHz
- 40 dBc Output impedance: <3 ohm at DC, <10 ohm at
1 MHz (load impedance 500 ohm, 500 pF, max. output current 40
mA peak) DC offset range: four times the specified range
of the standard instrument
VXI Characteristics
VXI device type: Message-based Data transfer
bus: A16/24, D16 Master, A16/D16 Slave Size: C Slots:
1 Connectors: P1/2 Shared memory: n/a VXI busses:
TTL Trigger Bus (T) C-size compatibility: Yes
Command module firmware: n/a Command
module firmware rev: n/a I-SCPI Win 3.1: n/a I-SCPI
Series 700: n/a C-SCPI LynxOS: n/a C-SCPI Series 700:
n/a HP VEE Drivers: n/a VXIplug&play Win Framework:
Yes VXIplug&play Win95/NT Framework: Yes VXIplug&play
HP-UX
Ordering Information:
Agilent HP E1440A VXI 21 MHz Synthesized Function/Sweep
Generator
| Agilent E1440A |
VXI 21 MHz Synthesized Function/Sweep Generator |
| Opt 001 |
High voltage output |
| Opt 0B1 |
Add manual set |
| Opt 0B2 |
Operation manual |
| Opt 0B3 |
Service manual |
|