California Instruments 2750L-BTS Ballast Test System, 3000VA

The L Series consist of a range of AC Power Source products with output power levels from 833 VA to 30 kVA. Using high frequency PWM technology, these units provide precision AC power in single, dual (split) or three phase configurations.
All L Series units use dual voltage ranges and programmable current limits to protect the unit under test. Standard voltage ranges of 135 and 270 Volt rms are typical but optional output range options offer voltage ranges up to 400 V rms. The complete solid state, high switching speed PWM architecture of the L Series ensures low noise and low distortion output over a wide range of power levels.

Controller Capabilities
The L Series uses a modular plug-in controller approach (except for -HGA controller type) that enables several controllers with different feature levels to be selected. For simple manual operation, the -M controller provides rotary knobs for setting voltage and frequency. Both front panel and computer control is available using one of the -P, -PT or -HGA controller options. Units equipped with -P, -PT or -HGA controller options also offer built in measurement functions. The following controller options are available for the L Series.

Controller type Frequency range* Special features
-M 45 Hz - 5000 Hz or
45 Hz - 2000 Hz
Manual controls
-P 45 Hz - 5000 Hz ± 0.05 % FS RMS load regulation
Transient generation
-PT 45 Hz - 550 Hz Fast settling time, average
response load regulation.
Transient generation
-HGA
(external 3.5 inch height chassis,
not a plug in)
Note: Not available for all L Series
models. Contact sales department
for configuration details.
45 Hz - 66 Hz Arbitrary waveform
generation and harmonic
analysis

All programmable controllers offer output measurement functions for voltage, rms current, power, power factor, phase (three phase systems) and frequency. Three phase models can be used in single phase mode by adding the -MODE option which allows automatic switching between single and three phase modes from both the front panel and the IEEE-488 bus.
Programmable controllers also include a standard IEEE-488 remote control interface and a Windows ™ Graphical User Interface (GUI) for PC control. The -HGA controller offers both IEEE-488 and RS232C and is supplied with a Windows ™ GUI program as well. This saves the end user from having to develop his own program and makes routine tasks like running product test sequences simple.

Features

á Single phase, three phase and split phase configurations
á Mode option allows switching between single and three phase output
á Output frequency range up to 5000 Hz.
á Full output VA with 0 to 1 power factor
á High peak current capability with up to 9:1 crest factor
á Panel height between 5.25" and 31.5" based on power level
á Efficiency 75% typical or better. Generates less heat and consumes less power
á Abbreviated Plain English (APE) programming language support and free Windows ™ application software
á Optional MIL-STD704 and RTCA-DO160C test firmware

Applications
The extended output frequency range of 17 Hz to 5000 Hz (-P Controller only with certain models) makes the L Series suitable for both commercial and avionics applications. Dual voltage ranges provide maximum output current at any programmed voltage up to 400 Volt. (-EHV option, some options are mutually exclusive) The built in measurement functions often eliminate the need for additional test equipment. Due to the many configuration possibilities of the L Series family of products, California Instruments is able to supply systems that specifically meet your application requirements. The L Series offers many options and configuration choices, some of which affect output ranges. Contact factory for configuration information.

Remote Control
All L Series AC sources shipped with a programmable controller (-P, -PT or -HGA) include a Windows ™ Graphical User Interface (GUI) program. The GUI supports all functions and capabilities of the L Series controller used. The following tasks can easily be performed through this graphical user interface using either the IEEE-488 or RS232C (HGA controller only) interface:
á Control all output parameters such as voltage, current limit, phase and frequency.
á Compile lists of transient programs on disk for quick recall and execution
á Measure and record key output parameters such as volt rms, current, peak current, real power, and power factor.
á Run RTCA/DO160C (requires option -160) or MIL-STD 704D tests (requires option -704)
á Define custom waveforms by specifying their harmonic content (HGA only)
á Define custom waveforms by free-hand drawing with the mouse (HGA only)
á Analyze harmonic content of both voltage and current (HGA only)
á Monitor remote control commands over the bus using the built in command viewer window to quickly learn how to program the L Series yourself.

Of course, drivers for popular programming environments such as LabView ™ are available as well for custom software development. If needed, the front panel can be locked out to prevent operator intervention during test runs.

Specifications

Specification Standard Programmable
Type -P
Transient Programmable
Type -PT
Manual Control
Type -M
Voltage
Ranges
Standard 0 - 135 V or 0 - 270 V.
All voltages are L-N.
Programmable range change included.
Program different amplitudes each phase.
Standard 0 - 135 V or 0 - 270 V.
All voltages are L-N.
Programmable range change included.
Program different amplitudes each phase.
Variable POT control from 0 to FS.
Optional (-RPV) 0 to FS based on 0 - 10 VDC input.
Switchable range change included.
Voltage
Accuracy
TRMS Sense:
±0.135 V from 5 V to 135 V
±0.54 V from 135 V to 270 V
All at 25¡C ±1¡C
(based on TRMS reading).
±0.7% FS from 5% FS to FS
Constant line, load and temperature at 25¡C ±1¡C.
Variable control.
(Analog meter indication)
Load
Regulation
TRMS sense: ±0.05% FS,
no load to full load.
-0.5% FS: 45 Hz - 100 Hz
-2.0% FS: 100 Hz - 550 Hz
±0.05%: 45 Hz to 5 kHz
Line
Regulation
±0.02% FS for ±10% line change ±2% of full output for a ±10% line change ±0.02% of full output for a ±10% line change
Stability
(Amplitude)
±0.015% FS over 24 hours at constant line, load and temperature. ±0.25% FS over 24 hours at constant line, load and temperature. ±0.015% FS over 24 hours at constant line, load and temperature.
Default Voltage 0.0/5.0 volts (field selectable) 0 volts N/A
Frequency
Range*
2750L-1P, 4500L-1P & all multi-box systems: 45 Hz to 2 kHz
-3P & 751L/1501L/20010L: 45 Hz to 5 kHz
6000L-1P, 12000L-1P: 45 Hz to 440 Hz
All models except 6000L and 12000L: 45 to 550 Hz
6000L, 12000L: 45 to 440 Hz
2750L-1M, 4500-1M & all multi-box systems: 45 Hz to 2 kHz
-3M & 751L/1501L/20010L: 45 Hz to 5 kHz
6000L, 12000L: 45 Hz to 440 Hz
Frequency
Resolution
0.01 Hz: 45 Hz to 99.99 Hz
0.1 Hz: 100 Hz to 999.9 Hz
1 Hz: 1000 Hz to 5000 Hz
0.01 Hz: 45 Hz to 99.99 Hz
0.1 Hz: 100 Hz to 550 Hz
3 digit resolution
Frequency
Accuracy
±0.001% of programmed value ±0.001% of programmed value ±0.005% of set value
Default
Frequency
(Initial Value)
Any (within range) Any (within range) Setting
Ext. Sync Input TTL level TTL level N/A
External
Amplitude
Modulation
N/A 0 to 5 V RMS signal generates 0 to 11% amplitude modulation of output voltage.
45 Hz to 5 kHz signal
0 to 5 V RMS signal generates 0 to 11% amplitude modulation of output voltage.
Phase Angle
Programming
Phases B and/or C relative to Phase A: 0 to ±360.0¡ in 0.1¡ increments Phases B and/or C relative to Phase A: 0 to ±360.0¡ in 0.1¡ increments N/A
Phase
Accuracy
±2¡ ±2¡ ±3¡
Distortion
(45 Hz - 550 Hz)
N/A 0 to 20% THD with sine waveform and 1% resolution N/A
Settling Time 16 mSec no-load,from 5V to 2% of final value;
16 mSec, full-load,to 15% of final value.
0.5 mSec N/A
Current Limit Adjustable trip Adjustable trip
Adjustable hold-off
Adjustable fold back with recovery
Remote On/Off Contact closure turn off Contact closure trip off Contact closure turn off

 

Model Total Power
35¡ C 1
Output
(Phase) 2
Total Current Rating on 135 V RNG 3 pH
A rms/pH
Size/Weight 6
A rms A peak 3 A peak 4
751L 833 VA 1 6.1 55.6 6015 n/a A
1501L 1,667 VA 1 12.3 55.6 6030 n/a A
1503L 1,667 VA 3 12.3 27.8 30 4.1 A
2001L 2,000 VA 1 14.8 55.6 60 n/a A
2750L 3,000 VA 1/3 22.2 83.3 90 7.4 B
4500L 5,000 VA 1/3 37 83.3 90 12.3 B
6000L 6,000 VA 1/3 44.4 157.4 170 14.8 B
9000L 10,000 VA 1/3 74 166.6 180 24.7 C
12000L 12,000 VA 1/3 90 314.8 340 30 C
13500L 15,000 VA 5 1/3 111.2 250.0 270 37 D
18000L 18,000 VA 5 1/3 133.2 472.0 510 44.4 D

Notes

(1) Derate power by 10% for operation at 50¡ C ambient or when using with -UP option.
(2) 1¿/3¿ systems are factory configured unless "MODE" option is specified.
(3) Repetitive peak current capability
(4) Peak in-rush drive capability (non-repetitive, less than 20 ms)
(5) Similar products available to 72,000 VA from the FCS Series. Contact factory for more information.
(6) See Dimensions, weight and environmental Table below for specifications.

Input

Voltage Input range must be specified when ordering
751L, 1501L, 1503L, 2001L 103 - 127 VAC, L-N, Single phase or
207 - 251 VAC, L-N, Single phase
2750L, 4500L, 6000L, 9000L,
12000L 13500L, 18000L
187 - 252 VAC, L-L, Three phase
(or 187 - 252 VAC, 1 Phase for 2750L only)
2750L, 4500L, 9000L, 13500L 342 - 456 VAC, L-L, Three phase with -UP option
Line Frequency 47 - 440 Hz
Hold up time > 10 msec

Standard Measurements, -P and -PT controllers

Parameter Range Accuracy* (±) Resolution*
AC Measurements
Frequency 16.00 - 99.99 Hz ± 0.02 Hz 0.01 Hz
100.0 - 500.0 Hz ± 0.2 Hz 0.1 Hz
500.0 - 999.9 Hz ± 0.5 Hz 0.1 Hz
1000 - 5000 Hz ± 10 Hz 1 Hz
Phase 0.0-360.0 ± 2¡ 0.1¡
RMS Voltage 0 -400 V 1 V 0.1 V
RMS Current 0 - 40 A 0.1 A 0.01 A
0 - 400 A 1 A 0.1 A
Real Power 0 - 2 kW 10 W 1 W
>0 - 20 kW 100 W 10 W
Apparent Power 0 - 2 kVA 10 VA 1 VA
0 - 20 kVA 100 VA 10 VA
Power Factor 0.000 - 1.000 0.001

System

Waveforms
Waveform Types Sine
Square (optional on -P and -PT)
Clipped Sine, 0 - 20 % THD
User defined (-HGA controller only)
User defined waveform storage -HGA 18 user defined arbitrary waveforms of 4096 points.
Transient Programming
Transient Types
Voltage: drop, step, sag, surge, sweep
Frequency: step, sag, surge, sweep
Voltage and Frequency: step, sweep
Transient register Parameters: Voltage, Frequency, Time, Slew rate, , Phase angle
Transient storage -P and -PT : up to 16 transient registers
-HGA: up to 32 transient steps per list
Time resolution 1 msec
Time range -P and -PT : 1 msec - 9999 sec
-HGA : 1 msec - 300 sec

Connectors

Line input Input terminal block
Output Output terminal block
Interfaces -P and -PT : 24 pin IEEE-488
-HGA : 24 pin IEEE-488 and 9 pin RS232C
Remote sense Remote voltage sense terminal block
Miscellaneous System interface
Clock and Lock BNC (2) (for rack enclosed system use only)

Dimensions, weight, environmental

A B C D
Height 5.25" / 133 mm 10.5" / 267 mm 21" / 533 mm 31.5" 800 mm
Width 19" / 483 mm 19" / 483 mm 19" / 483 mm 19" / 483 mm
Depth 23" / 584 mm 23" / 584 mm 23" / 584 mm 23" / 584 mm
Weight (net) 85 lbs / 39 Kg 180 lbs / 82 Kg 360 lbs / 164 Kg 540 / 250 Kg
Vibration and Shock designed to meet NSTA project 1A transportation levels
Cooling Forced air cooling, side air intake, rear exhaust.
Operating humidity 0 to 95 % RH, non condensing
Operating Altitude 0 - 6000 feet
Operating Temperature 0 to 50 ¡ Celsius
Storage Temperature -40 to + 85 ¡ Celsius

 

 


 
 
  Company    |   Contact Us   |   Site Map 1.800.417.4370
 
 
 
 