Home   |   Contact Us   |   Shopping Cart   |   Quote Cart  0 items   |   Login
Electronic Test Equipment - MetricTest   Advanced
1.800.417.4370
 
 

Fluke 5520A Calibrator

The Fluke 5520A builds on the 5500A's capabilities, extending its workload coverage even further. Its improved accuracy, expanded ranges, and capability to calibrate 5 1/2 and 6 1/2 digit multimeters can cover virtually all your high-performance workload, including:

  • Handheld and bench meters (analog and digital) up to 6 1/2 digits
  • Thermocouple and RTD thermometers
  • Process calibrators
  • Data loggers
  • Strip and chart recorders
  • Watt meters
  • Power harmonics analyzers
  • Current clamps and clamp meters
  • Analog or digital handheld and bench oscilloscopes (with options)
  • Panel meters
  • Graphical multimeters

Instead of buying several standards to calibrate a variety of instruments, Fluke multi-product calibrators can meet your needs with one compact affordable tool.

Features

The Fluke 5520A Multi-Product Calibrator sources direct voltage and current, alternating voltage and current with multiple waveforms and harmonics, two simultaneous voltage outputs or voltage and current to simulate dc and ac power with phase control, resistance, capacitance, thermocouples and RTDs. The 5520A can also measure thermocouple temperature, humidity (with probe) and pressure using one of 28 Fluke 700 Series pressure modules. Three options add the capability to calibrate oscilloscopes up to 1.1 GHz.

Wide workload coverage

Because it can calibrate so much of your workload, the 5520A makes your calibration investment go farther. Instead of several instruments to purchase, maintain and transport, the 5520A combines all the capability in one compact, portable and affordable tool.

Though it can do a lot, the 5520A is simple to use. Its front panel is logically arranged and clearly labeled. And its bright LCD displays provide all the information you need at a glance. The 5520A is so intuitively easy to use, you'll seldom need to consult the manual. For almost every operation, your hand moves naturally from left to right. Values are entered on the calculator-style keyboard. Output values can be stepped up or down easily in even decade multiples (1-2-5 sequence in scope mode) using the multiply and divide keys. You can even view the uncertainty for any value at the press of a key

The basic 5520A can tackle a tremendous amount of work. And as your workload changes, your 5520A can adapt. Three options are available to address the calibration requirements of oscilloscopes. If your workload is made up mostly of oscilloscopes up to 300 MHz bandwidth, the 5500A-SC300 option is an affordable solution. If your workload extends to 600 MHz, the 5500A-SC600 adds complete, powerful oscilloscope calibration capability to your 5520A. The 5520A-SC1100 option enables the 5520A to calibrate all of your oscilloscopes up to 1.1 GHz with rise times to 300 ps - even 125 ps with the 5800A/TDP tunnel diode pulser option. Any of these options can be fitted at time of purchase, or added later as your needs change.

If three-phase power measurement tools are part of your workload, the 5520A can greatly simplify the task. Its phase locking capability makes it possible to "lock" three 5520As together to simulate three phase power up to nearly 21 kW with multiple waveforms, harmonics and precise phase control. Using this capability, you can also "sum" the current outputs of multiple 5520As to source 100A or more to directly verify current clamps or power transformers.

The 5520A isn't limited to the controlled environment of the calibration lab. It's made to go where the work is. It's compact so it's easy to take it where the work is. Its rugged construction means you don't have to treat it like a delicate lab instrument. And it's self-contained. No additional test pad is required.

Every 5520A is shipped with a certificate of calibration with results data, traceable to nationally recognized standards. When it comes time to recalibrate your instrument, Fluke service facilities are available worldwide to ensure the work is done right and quickly.

With quality management standards like ISO 9000, there is a lot more to calibration than just making measurements. You also have documentation, control and reporting requirements to meet.

Our optional Microsoft Windows¨-based 5500/CAL simplifies the documentation of your procedures, adequacy and traceability as required by ISO 9000 and other similar quality standards. It also collects and reports calibration results information and helps you to consistently, quickly and efficiently calibrate a wide variety of instruments. With it your entire calibration process Ñ from creating and executing procedures through results data collection and reporting û can be automated. Thousands of procedures û and the list is growing û are included at no extra charge. They can be run as is or used as the basis to create your own new procedures.

5500/CAL is part of a suite of applications called MET/CAL¨ Plus. 5500/CAL is designed around RS-232 instrument control and operates with the 5500A, 5520A, and 5800A calibrators. If you need IEEE instrument control or want to automate the calibration of additional instruments, you can step up to MET/CAL¨. And you can add test and measurement asset management capabilities with MET/TRACK.

All three applications are driven by a powerful, industry standard SQL data base server, SQL Anywhere¨ from Sybase, providing secure, dependable storage for your calibration information. And each application is completely modular so you can set up a single application on a personal computer or take advantage of its client/server design over your network.

Specifications *

DC Voltage

Range: 0 to ±1020V
Best 1 Year Specification: ±12 ppm of setting

DC Current

Range: 0 to ±20.5A
Best 1 Year Specification: ±100 ppm of setting

Resistance

Range: 0 to 1100 MW
Best 1 Year Specification: ±28 ppm of setting

AC Voltage

Range: 1 mV to 1020V
10 Hz to 500 kHz, Sine
Best 1 Year Specification: ±120 ppm of setting

AC Current

Range: 29 mA to 20.5A
10 Hz to 30 kHz, Sine
Best 1 Year Specification: ±0.06% of setting

Capacitance

Range: 0.19 nF to 110 mF
Best 1 Year Specification: ±0.25% of setting

Thermocouple Source and Measure

Range: -250¡C to +2316¡C (11 types)
Best 1 Year Specification: ±0.14¡C of setting

RTD Source

Range: -200 to +630¡C (8 types)
Best 1 Year Specification: ±0.03¡C of setting

DC Power

Range: 10.9 mW to 20.5 kW
Best 1 Year Specification: ±0.023% of setting

AC Power

Range: 10.9 mW to 20.5 kW
45-65 Hz, PF=1
Best 1 Year Specification: ±0.08% of setting

Phase

Range: 0 to ± 179.99¡
10 Hz to 30 kHz
Best 1 Year Specification: ±0.07¡ of setting

Frequency

Range: 0.01 Hz to 2.0 MHz
Best 1 Year Specification: ±2.5 ppm of setting

 

General Specifications *

Warm up Time

Twice the time since last warmed up, to a maximum of 30 minutes

Settling Time

Less than 5 seconds for all functions and ranges except as noted

Standard Interfaces

IEEE-488 (GPIB), RS-232, 5725A (5500A only)

Temperature Performance

Operating: 0¡C to 50¡C
Calibration (tcal): 15¡C to 35¡C
Storage: -20¡C to 70¡C

Temperature Coefficient

Temperature coefficient for temperatures outside tcal ± 5¡C is 10% of the 90-day specification (or 1-year, as applicable) per ¡C

Relative Humidity

Operating: <80% to 30¡C, <70% to 40¡C, <40% to 50¡C
Storage: <95%, non-condensing

Altitude

Operating: 3,050m (10,000 ft) maximum
Non-operating: 12,200m (40,000 ft) maximum

Safety

Designed to comply with IEC 1010-1 (1992-1); ANSI/ISA-S82.01-1994; CAN/CSA-C22.2 No. 1010.1-92

Analog Low Isolation

20V

EMC

5520A: Designed to comply with IEC 61326-1/1997
5500A: Designed to comply with FCC Rules Part 15

Line Power

Line Voltage (selectable): 100V, 120V, 220V, 240V
Line Frequency: 47 to 63 Hz
Line Voltage Variation: + 10% about line voltage setting

Power Consumption

5520A: 600 VA
5500A: 300 VA

Dimensions

Height: 17.8 cm (7 in), standard rack increment, plus 1.5 cm (0.6 in) for feet on bottom of unit
Width: 43.2 cm (17 in), standard rack width
Depth: 47.3 cm (18.6 in) overall

Weight (without options)

5520A: 22 kg (49 lbs.)
5500A: 20 kg (44 lbs.)

Absolute Uncertainty Definition

5500A and 5520A uncertainty specifications include stability, temperature coefficient, linearity, line and load regulation, and the traceability of the external standards used for calibration. You do not need to add anything to determine the total uncertainty of your calibrator for the temperature range indicated.

Specification Confidence Interval

> 99%



 


 
 
  Company    |   Contact Us   |   Site Map 1.800.417.4370
 
 
 
 