Fluke 5700A-03 Calibrator
The 5700A and 5720A are five-function calibrators designed to address
the most demanding electrical calibration workload. They source direct
and alternating voltage and current and resistance. A wideband voltage
option extends overall ac bandwidth to 10 Hz to 30 MHz to cover to RF
voltmeters. Both are compatible with the 5725A, 5220A and 5205A/5215A
amplifiers.
Features
In service since 1988, the 5700A has undergone continuous
improvements to become the 5700A Series II, one of the most tested and
reliable high precision calibrators Fluke has ever produced. Considered
the calibration standard worldwide, the 5700A delivers high value as well
as accuracy covering 5 1/2 to 7 1/2 digit dmms. Plus it offers the same
ease of use, low cost of ownership, rugged design, simplified support
and confidence building features as the 5720A.
Specifications for both the 5700A and 5720A are stated with
a choice of confidence levels. Now, you can use the conservative 99% specifications
that Fluke traditionally publishes, or more aggressive 95% specifications
recommended in many international procedures. In addition to allowing
you to trade a small amount of risk for better performance, 95% confidence
level specifications allow easier measurement intercomparisons. Both specifications
are available at a press of the SPEC key for any output.
Both the 5700A and 5720A can emulate, via the remote interface,
either a 5100B or 5200A calibrator, permitting them to replace those older
calibrators in automated systems with little or no impact on software.
In addition, it is compatible with the 5725A Amplifier, 5220A Transconductance
Amplifier and 5205A or 5215A Precision Power Amplifiers.
Both the 5700A and the 5720A feature Artifact Calibration.
Only three artifact standards _ a 10V dc reference and 1W and 10 kW resistance
references are required to calibrate all ranges and functions to full
specifications. Front panel instructions prompt the operator to make connections
and inputs each step of the way. The calibrator controls the process,
which takes only about an hour, compared to several hours using traditional
methods.
During the process, assigned values of an external artifact
are transferred to a large array of multidimensional parameters within
the 5720A. The calibrator takes over the manual metrology functions of
establishing ratios and making comparisons, as well as controlling the
measurement process.
To assure confidence, the 5700A and 5720A can check themselves
against their own internal standards to assure everything is working as
expected. Those results can be printed out or downloaded to a computer.
Thousands of 5700A calibrators in service around the world
prove Artifact Calibration delivers fast, easy and inexpensive calibration
along with the confidence that your instrument is performing as expected
between calibrations.
Specifications
|
Specifications
|
|
DC Voltage
|
Ranges: 0 to ±1100V
5700A Best 1 Year Specification: ±7.0 ppm of setting
5720A Best 1 Year Specification: ±3.5 ppm of setting
|
|
AC Voltage
|
Ranges:
220 mV to 1100V
10 Hz to 1 MHz
5700A Best 1 Year Specification: ±75.0 ppm of setting
5720A Best 1 Year Specification: ±45.0 ppm of setting
|
|
Resistance
|
Ranges: 0 to 100 MW, 18 values in x1 and x1.9
5700A Best 1 Year Specification: ±12.0 ppm
5720A Best 1 Year Specification: ±8.5 ppm
|
|
DC Current
|
Ranges: 0 to Û2.2A (11A with 5725A)
5700A Best 1 Year Specification: Û50 ppm of setting
5720A Best 1 Year Specification: Û35 ppm of setting
|
|
AC Current
|
Ranges:9 mA to 2.2A (11A with 5725A) 10 Hz to 10 kHz
5700A Best 1 Year Specification: ±140 ppm of setting
5720A Best 1 Year Specification: ±120 ppm of setting
|
|
Wideband AC Voltage Option
|
Ranges:
300 mV to 3.5V
10 Hz to 30 MHz
5700A Best 1 Year Specification: ±0.4% of setting
5720A Best 1 Year Specification: ±0.4% of setting
|
|
General Specifications *
|
|
Warm-up time
|
2x the time since last warmed up, to a maximum of
30 minutes
|
|
System installation
|
Rear output configuration and rack-mount kit available
|
|
Standard interfaces
|
IEEE-488, RS-232, 5725A, 5205A or 5215A, 5220A, phase
lock in (BNC), phase reference out (BNC)
|
|
Temperature performance
|
Operating: 0¡C to 50¡C
Calibration: 15¡C to 35¡C
Storage: -40¡C to 75¡C
|
|
Relative humidity
|
Operating: < 80% to 30¡C, < 70% to 40¡C, < 40% to
50¡C
Storage: < 95%, non-condensing. A power-on stabilization period
of four days may be required after extended storage at high temperature
and humidity.
|
|
Safety
|
Designed to comply with UL311; IEC 348-1978; IEC 66E
(CO)4; CSA 556B
|
|
Guard isolation
|
20 volts
|
|
EMI/RFI
|
Designed to comply with FCC Rules Part 15,
Subpart J, Class B; VDE 0871, Class B
|
|
Line power
|
47 to 63 Hz; + 10% allowed about selectable nominal
line voltage: 100V, 110V, 115V, 120V, 200V, 220V, 230V, 240V
Maximum power: 5700A/5720A: 300 VA; 5725A, 750 VA
|
|
Size
|
5700A/5720A: Height 17.8 cm (7 in), standard rack
increment, plus 1.5 cm (0.6 in) for feet; width 43.2 cm (17 in),
standard rack width; depth 63.0 cm (24.8 in) overall; 57.8 cm (22.7
in), rack depth
5725A: Height 13.3 cm (5.25 in); width and depth same as 5700A/5720A
Both units project 5.2 cm, (2 in) from rack front
|
|
Weight
|
5700A/5720A: 27 kg (62 lbs.)