Home   |   Contact Us   |   Shopping Cart   |   Quote Cart  0 items   |   Login
Electronic Test Equipment - MetricTest   Advanced
1.800.417.4370
 
 

Fluke 5700A Calibrator


The 5700A and 5720A are five-function calibrators designed to address the most demanding electrical calibration workload. They source direct and alternating voltage and current and resistance. A wideband voltage option extends overall ac bandwidth to 10 Hz to 30 MHz to cover to RF voltmeters. Both are compatible with the 5725A, 5220A and 5205A/5215A amplifiers.

Features

In service since 1988, the 5700A has undergone continuous improvements to become the 5700A Series II, one of the most tested and reliable high precision calibrators Fluke has ever produced. Considered the calibration standard worldwide, the 5700A delivers high value as well as accuracy covering 5 1/2 to 7 1/2 digit dmms. Plus it offers the same ease of use, low cost of ownership, rugged design, simplified support and confidence building features as the 5720A.

Specifications for both the 5700A and 5720A are stated with a choice of confidence levels. Now, you can use the conservative 99% specifications that Fluke traditionally publishes, or more aggressive 95% specifications recommended in many international procedures. In addition to allowing you to trade a small amount of risk for better performance, 95% confidence level specifications allow easier measurement intercomparisons. Both specifications are available at a press of the SPEC key for any output.

Both the 5700A and 5720A can emulate, via the remote interface, either a 5100B or 5200A calibrator, permitting them to replace those older calibrators in automated systems with little or no impact on software. In addition, it is compatible with the 5725A Amplifier, 5220A Transconductance Amplifier and 5205A or 5215A Precision Power Amplifiers.

Both the 5700A and the 5720A feature Artifact Calibration. Only three artifact standards _ a 10V dc reference and 1W and 10 kW resistance references are required to calibrate all ranges and functions to full specifications. Front panel instructions prompt the operator to make connections and inputs each step of the way. The calibrator controls the process, which takes only about an hour, compared to several hours using traditional methods.

During the process, assigned values of an external artifact are transferred to a large array of multidimensional parameters within the 5720A. The calibrator takes over the manual metrology functions of establishing ratios and making comparisons, as well as controlling the measurement process.

To assure confidence, the 5700A and 5720A can check themselves against their own internal standards to assure everything is working as expected. Those results can be printed out or downloaded to a computer.

Thousands of 5700A calibrators in service around the world prove Artifact Calibration delivers fast, easy and inexpensive calibration along with the confidence that your instrument is performing as expected between calibrations.

Specifications

Specifications

DC Voltage

Ranges: 0 to ±1100V
5700A Best 1 Year Specification: ±7.0 ppm of setting
5720A Best 1 Year Specification: ±3.5 ppm of setting

AC Voltage

Ranges:
220 mV to 1100V
10 Hz to 1 MHz
5700A Best 1 Year Specification: ±75.0 ppm of setting
5720A Best 1 Year Specification: ±45.0 ppm of setting

Resistance

Ranges: 0 to 100 MW, 18 values in x1 and x1.9
5700A Best 1 Year Specification: ±12.0 ppm
5720A Best 1 Year Specification: ±8.5 ppm

DC Current

Ranges: 0 to Û2.2A (11A with 5725A)
5700A Best 1 Year Specification: Û50 ppm of setting
5720A Best 1 Year Specification: Û35 ppm of setting

AC Current

Ranges:9 mA to 2.2A (11A with 5725A) 10 Hz to 10 kHz
5700A Best 1 Year Specification: ±140 ppm of setting
5720A Best 1 Year Specification: ±120 ppm of setting

Wideband AC Voltage Option

Ranges:
300 mV to 3.5V
10 Hz to 30 MHz
5700A Best 1 Year Specification: ±0.4% of setting
5720A Best 1 Year Specification: ±0.4% of setting

 

General Specifications *

Warm-up time

2x the time since last warmed up, to a maximum of 30 minutes

System installation

Rear output configuration and rack-mount kit available

Standard interfaces

IEEE-488, RS-232, 5725A, 5205A or 5215A, 5220A, phase lock in (BNC), phase reference out (BNC)

Temperature performance

Operating: 0¡C to 50¡C
Calibration: 15¡C to 35¡C
Storage: -40¡C to 75¡C

Relative humidity

Operating: < 80% to 30¡C, < 70% to 40¡C, < 40% to 50¡C
Storage: < 95%, non-condensing. A power-on stabilization period of four days may be required after extended storage at high temperature and humidity.

Safety

Designed to comply with UL311; IEC 348-1978; IEC 66E (CO)4; CSA 556B

Guard isolation

20 volts

EMI/RFI

Designed to comply with FCC Rules Part 15,
Subpart J, Class B; VDE 0871, Class B

Line power

47 to 63 Hz; + 10% allowed about selectable nominal line voltage: 100V, 110V, 115V, 120V, 200V, 220V, 230V, 240V
Maximum power: 5700A/5720A: 300 VA; 5725A, 750 VA

Size

5700A/5720A: Height 17.8 cm (7 in), standard rack increment, plus 1.5 cm (0.6 in) for feet; width 43.2 cm (17 in), standard rack width; depth 63.0 cm (24.8 in) overall; 57.8 cm (22.7 in), rack depth
5725A: Height 13.3 cm (5.25 in); width and depth same as 5700A/5720A
Both units project 5.2 cm, (2 in) from rack front

Weight

5700A/5720A: 27 kg (62 lbs.)
5725A: 32 kg (70 lbs.)


 


 
 