Keithley 236 High Voltage Source/Measure

 

 Features

The 236, 237 and 238 Source-Measure Units (SMU) are fully programmable instruments, capable of sourcing and measuring voltage or current simultaneously. These systems are really four instruments in one: voltage source, current source, voltage measure and current measure.


The 236 will source voltage from 100µV to 110V, and current from 100fA to 100mA. It can also measure voltage from 10µV to 110V and current from 10fA to 100mA. The 237 offers the same capabilities with a decade enhancement in voltage source and measure (1100V). In this higher voltage range, current source and measure is 10mA maximum. The 238 offers a decade enhancement in current source and measure (1A). In this higher current range, voltage source and measure is 15V maximum.


The 236, 237 and 238 will measure very small currents and voltages. With current sensitivity of 10fA, measurement capabilities are equal to those of an electrometer. Selectable integration and the filtering of multiple measurements enhances sensitivity for demanding applications

 

 Specifications

VOLTAGE SOURCE V
  RANGE
(Max. Value)
STEP SIZE ACCURACY
(1 Year,
18°-28°)
       
236, 237 ±1.1000 V2 100µV ±(0.033%+650 µV .
+ [IO/IFS] x 450µV)
  ±11.000 V 1mV ±(0.033% + 2.4mV)
  ±110.00 V 10mV ±(0.033% +  24mV)
237 Only ±1100.0 V 100mV ±(0.04 % + 240mV)
238 Only ±1.5000 V 100µV ±(0.033% +800 µV
+ [IO/IFS]x 600µV)3
  ±15.000 V 1mV ±(0.033% + 2.7mV)
  ±110.00 V 10mV ±(0.033% + 24mV)

 

VOLTAGE MEASURE V
  RESOLUTION ACCURACY
(1 Year,
18°-28°)
  4-Digit 5-Digit  
236, 237 100µV 10µV ±(0.028% + 300µV
+ [IO/IFS] x 450µV)
  1mV 100µV ±(0.025%+  1mV)
  10mV 1mV ±(0.025%+ 10mV)
237 Only 100mV 10mV ±(0.035%+100mV)
238 Only 100µV 10µV ±(0.028% + 450µV
+ [IO/IFS] x 600µV)
  1mV 100µV ±(0.025%+  1mV)
  10mV 1mV ±(0.025%+ 10mV)

 

CURRENT SOURCE I
  RANGE
(Max. Value)
STEP
SIZE
ACCURACY
(1 Year,
18°-28°C)
       
All ±1.0000nA 100fA ±(0.3 %+ 450 fA)
  ±10.000nA 1pA ±(0.3 %+   2 pA)
  ±100.00nA 10pA ±(0.21%+  20 pA)
  ±1.0000µA 100pA ±(0.05%+ 200 pA)
  ±10.000µA 1nA ±(0.05%+   2 nA)
  ±100.00µA 10n ±(0.05%+  20 nA)
  ±1.0000mA 100nA ±(0.05%+ 200 nA)
  ±10.000mA 1µA ±(0.05%+   2 µA)
  ±100.00mA 10µA ±(0.1 %+  20 µA)
238 Only ±1.0000A 100µA ±(0.12%+ 700 nA)

 

CURRENT MEASURE I
  RESOLUTION ACCURACY
(1 Year,
18°-28°C)
 
  4-Digit 5-Digit  
All 100fA 10fA ±(0.3  % + 100fA)2
  1pA 100fA ±(0.3  % +   1pA)
  10pA 1pA ±(0.21 % + 6pA)
  100pA 10pA ±(0.04 % + 6pA)
  1nA 100pA ±(0.035% + 700pA)
  10nA 1nA ±(0.035% + 6 nA)
  100nA 10nA ±(0.035% + 60nA)
  1µA 100nA ±(0.038% + 600nA)
  10µA 1µA ±(0.1  % + 6µA)
238 Only 100µA 10µA ±(0.12 % + 300 µA)

 