LECROY LT344L Digital Oscilloscope

 

 Features

 

  • 500 MHz bandwidth
  • 1 GS/s max single-shot sample rate
  • 25 GS/s for repetitive signals and
  • Up to two million data points to view signals with high resolution over longer time intervals

 

 Specifications

Autosetup

Automatically sets timebase, trigger, and sensitivity to display a wide range of repetitive signals; vertical find automatically sets sensitivity for the selected input signal.

Acqusition Systems

Model Bandwidth Channels Sample Rate Acquisition Memory
LT332 500 MHz Two 200 MS/s 100 kpts/Ch
LT342 500 MHz Two 500 MS/s 250 kpts/Ch
LT342L 500 MHz Two 500 MS/s 1 Mpts/Ch
LT344 500 MHz Four 500 MS/s 250 kpts/Ch
LT344L 500 MHz Four 500 MS/s 1 Mpts/Ch
LT364 500 MHz Four 1 GS/s/2 Ch 500 kpts/Ch
LT364L 500 MHz Four 1 GS/s/2 Ch 2 Mpts/Ch
LT224 200 MHz Four 200 MS/s 100 kpts/Ch

Waverunner Triggers

Name Description Edge or SMART Trigger®
Edge/Slope/Window/Line Triggers when signal meets slope and level condition. Edge
Signal pulse width Triggers on glitches down to 2 ns. Width selectable from <2.5 ns to 20 s or on intermittent faults. SMART
Signal interval Triggers on intervals selectable between 10 ns and 20 s. SMART
State or edge qualified Triggers on input only after a defined state or edge occurred on another channel. Delay between sources is selectable by time or events. SMART
Dropout Triggers if the input signal drops out for longer than selected time between 25 ns and 20 s. SMART
Pattern* Triggers on logic combination of five inputs - CH1, CH2, CH3, CH4, and EXT trigger, where each source can be defined as High, Low, or Don't Care. SMART
TV Triggers on line (up to 1500), NTSC, or nonstandard video, choice of fields**. SMART

* LT364/L only

** 1, 2, 4, 8 fields for LT364/L; odd or even fields for all other models

Rapid Signal Processing

Memory:
16 Mbytes: LT364, LT344, LT342, LT322, LT224
32 Mbytes: LT364L, LT344L, LT342L

Internal Waveform Memory

Zoom and Math: A, B, C, D

**Store full-length waveforms with 16 bits/data point.

Setup Storage

For front panel and instrument status: Four non-volatile memories and floppy drive are standard. PC card hard drive and memory card are optional.

Math Tools

Simultaneously perform up to four math processing functions; traces can be chained together to perform math on math. standard math tools:

Standard Math Tools:

average (summed to 4000 sweepts)
difference
enhanced resolution (to 11 bits vertical)
envelope
FFT of 50 kpoint waveforms
floor
identity
negate
product
ratio
reciprocal (invert)
resample (deskew)
rescale (with units)
roof
sin x/x
sum

 

Measure Tools

Automated Measurements: Display any five parameters together with their average, high, low, and standard deviations.

Standard Measure Tools:

amplitude duty cycle -overshoot
area fall 90-10% peak-to-peak
base fall 80-20% period
cycle mean frequency phase
cycle rms maximum rise 10-90%
cycles mean rise 20-80%
data median rms
delay minimum top
delta delay +overshoot width

Pass/Fail

Test any five parameters against selectable thresholds. Limit testing is performed using masks created on the scope or PC. Setup a pass or fail condition to initiate actions such as hardcopy output, save waveform to memory, GPIB SRQ, or pulse out.

Extended Math and Measurements Option

Adds math and advanced measurements for general-purpose applications. Includes all standard math and measurement tools, plus:

Extended Math Tools:

absolute value log (base 10)
differentiate integrate
exp (base e) square
exp (base 10) square root
log (base e) trend (datalog)

Extended Measure Tools:

cycle medium fall @ level; % and volts
cycle std. deviation first point
delta time @ level; % and units last point
delta time @ level from trigger number of points
delta time from clock to data + (setup time) rise @ level; % and volts
delta time from clock to data - (hold time) std. deviation
duration

Waveanalyzer Option (WAVA)

Includes the Extended Math and Measure Tools as well as expanded capabilities for performing FFTs, averaging, histograms, and histogram parameters.

WaveAnalyzer Tools:

Histograms with 18 histogram parameters
Summed averaging to one million sweeps
Continuous weighted averaging
FFT (up to 1 Mpoint waveforms)
FFT power averaging
FFT power density, real, and imaginary

Other Application Solutions

Jitter and Timing Analysis (JTA), PowerMeasure System, Telecom Mask Testing

Interface 

Remote Control: Full control via GPIB and RS-232-C (standard) or via Ethernet for all instrument controls and internal functions (optional).

Ethernet Port (Optional): 10 Base-T, TCP/IP Protocol

Floppy Drive: Internal, DOS-format, 3.5" high-density

PC Card Slot (Optional): Supports memory and hard drive cards.

External Monitor Port: 15-pin D-Type VGA-compatible, 640 x 480 RGB

Centronics Port: Parallel printer interface

Internal Graphics Printer (Optional): Provides hardcopy output in <10 seconds.

Outputs

Calibrator Signal: 500 Hz–1 MHz square wave, -1.0 to +1.0, test point, and ground lug on front panel

Control Signals: Choice of trigger ready, trigger out, or pass/fail status; TTL levels into 1 MOhm at rear panel BNC (output resistance 300 Ohms ±10%)

General

Operating Conditions: Temperature 5–40°C; humidity 80% non-condensing at 40°C; altitude <=2,000 meters

Shock and Vibration: Conforms to MIL-PRF-28800F, Class 4

Power Requirements: 90-132 V AC and 180-250 V AC; 45-66 Hz; maximum power dissipation: 150 VA - 230 VA, depending on model

Certifications: CE, UL and cUL

Dimensions (HWD): 210mm x 350mm x 300mm; 8.3" x 13.8" x 11.8" (height excludes feet)

Weight: 8 kg; 18 lbs

 