Home > Rohde & Schwarz > NRVD

Rohde & Schwarz NRVD Dual-channel Power Meter

Characteristics

Measurement Functions - Average power, pulse power, peak envelope power. AM, reflection, DC voltage (depending on sensor).

Frequency and Level Range - DC to 40 GHz, 100 pW to 30 W, 9 kHz to 3 GHz, 200µV to 1000V (depending on sensor).

Probes and Sensors - All NRV sensors and URV5 probes.

Display - LCD for digits, units, menu-guided operation and analog display, adjustable backlighting.

Display of Results - Single-channel (with optional display of correction frequency) or dual-channel.

Absolute Readout - W, dBm, V, dBµV, dBV.

Relative Readout - dB, difference, percent and ratio, relative to a stored reference value or to the second measurement channel; VSWR, reflection coefficient, return loss in dB, AM modulation depth.

Analog Display - Automatic or with selectable scale.

Digital Display and Resolution - Max. 4.5 digits, resolution selectable (0.1/0.01/0.001 dB).

Display Filtering - Averaging over 1 to 512 readings to reduce display noise; manual or automatic setting depending on measurement range and resolution.

Display Noise - See sensors.

Measurement Rate - See table.

Accuracy of Power Readout in W (without sensors) -
0.4% (0.3%) + 1 digit (18 to 28°C)
0.9% (0.8%) + 1 digit (10 to 40°C)
1.4% (1.3%) + 1 digit (0 to 50°C).

Zero Adjustment - Manual or via IEC/IEEE bus, duration approx. 4 s.

Frequency Response Correction - Sensor-specific calibration data taken into account; numerical entry of test frequency (keyboard or via IEC/IEEE bus) or by frequency-proportional DC voltage.

Attenuation Compensation - External attenuation or gain taken into account; data entry via keyboard or IEC/IEEE bus, range ±200 dB.

Entry of Reference Value - Measured value on keystroke or numerical entry via keypad or IEC/IEEE bus.

Reference Impedance - For conversion between voltage and power, automatic readout of reference impedance from sensor data memory or numerical entry via keyboard or IEC/IEEE bus (for RF probe).

Remote Control - IEC 625 (IEEE 488), SCPI, control of all instrument functions.

Interface Functions - SH1, AH1, T6, L4, SR1, RL1, DC1, DT1, PP0, PP1.

DC Frequency Input -
NVRD: Opt. NVRD-B2.
Connector: BNC.
Input voltage range: ±12 V, linear with selectable scale.

DC Output -
NVRD: Opt. NVRD-B2.
Connector: BNC, Rout = 1 kiloohm, EMF proportional to analog display.
Left-/Right-hand Full-scale Value: Corresponding to 0/+3 V.
Accuracy: ±5 mV.
Channels: 1,2.

Input/Output Option NRVD-B2 - 2 simultaneous DC voltage outputs, DC frequency input, trigger input (TTL, active low).

Sensor Check Source

NRVD - Standard.

Output Power - 1 mW ±0.7%.

Frequency - 50 MHz.

VSWR - 1.05, <1.03.

RF Connector - N female.

General Data

Power Supply -
115 V + 15/-22%(-15%).
47 to 63 (440)  Hz;
230 V +  15/-22%
47 to 63 Hz, 13 VA
100/120/220 V ± 10%,
230 V -6/+15%;
47 to 400 Hz (25 VA).

Physical Characteristics

Dimensions

mm

in.

Height 147 5.8
Width 219 8.6
Depth 350 13.8

Weight

kg

lbs.

Net 4.5 9.9

Measurement Time in Seconds (from Trigger to Output of First Byte) Depending Upon Filter Setting.


Filter Number

Resolution

0 1 2 3 4 5 6 7 8 9 10 11 12
NRV-Z1 to -Z15 0.045 0.05 0.06 0.08 0.15 0.27 0.49 0.95 1.85 3.6 7.2 14.5 28.5
NRV-Z31 Mod. 02 1.04 1.04 1.05 1.07 1.13 1.24 1.44 1.84 2.7 4.3 7.5 14 27
NRV-Z31 to -Z33 Mod. 03, 04 0.135 0.14 0.15 0.17 0.23 0.34 0.54 0.94 1.77 3.4 6.6 13 26
NRV-Z32 Mod. 05 0.435 0.44 0.45 0.47 0.53 0.64 0.84 1.24 2.07 3.7 6.9 14 27
NRV-Z51 to -Z55 0.115 0.12 0.13 0.15 0.21 0.32 0.52 0.92 1.75 3.4 6.6 13 26
URV 5-Z2, -Z4, -Z7 0.065 0.07 0.08 0.1 0.2 0.38 0.72 1.45 2.8 5.5 11 22 44

 
 