Signal Acquisition System
Bandwidth - 200 MHz (TDS420A),
400 MHz (TDS430A, TDS460A).
Channels - 4 (2 on TDS430A).
Sample Rate - 100 MS/s on all
channels.
Sensitivity - 1 mV to 10 V/div
(with calibrated fine adjust).
Position Range - ±5 Divisions.
Offset Range - ±1 V
from 1 mV/div to 99.5 mV/div; ±10 V from 100 mV
to 995 mV/div; ±100 V from 1 V/div to 10 V/div.
DC Gain Accuracy - ±1.5%.
Vertical Resolution - 8-Bits (256
levels over 10.24 vertical divisions).
Analog Bandwidth Selections - 20 MHz,
100 MHz and full.
Input Coupling - AC, DC or GND.
Input Impedance Selections - 1 megaohm
in parallel with 15 pF or 50 Ohm (AC and DC coupling).
Maximum Input Voltage - 300 V
CAT II ±400 Vp. Derate at 20 dB/decade
above 1 MHz. 1 megaohm or GND coupled.
Channel Isolation - >100:1 at
100 MHz for any two channels.
AC Coupled Low Frequency Limit - <10 Hz
when AC 1 megaohm coupled. <200 kHz when AC
50 Ohm coupled.
Acquisition Modes
Peak Detect - High frequency and
random glitch capture. Captures glitches of 10 ns using acquisition
hardware at all real-time sampling rates.
Sample - Sample data only.
Envelope - Max/min values acquired
over one or more acquisitions, selectable from 2 to 2000, infinite.
Average - Waveform averages selectable
from 2 to 10,000.
Hi-res - Vertical resolution improvement
and noise reduction on low-frequency signals, e.g. 12-Bits at 10 ms/div
and slower. Enhanced vertical resolution (>12-Bits) for noise
reduction, on low frequency signals. Make precise low-level signal
measurements (up to 5 µV) with differential amplifier
(ADA400A).
Time Base System
Time Bases - Main, Delayed.
Time/Division Range - 1 ns
to 20 s/div.
Time Base Accuracy - 0.015% over
any interval >1 ms.
Record Length (Real Time and Equivalent
Time) - Sample points per channel: 500 to 120,000 points (with
Opt. XL).
Pre-trigger Position - Selectable
from 0 to 100% of record.
Triggering System
Triggers - Main, Delayed.
Main Trigger Modes - Auto, Normal,
Single Sequence.
Delayed Trigger - Delayed by time
or events.
Time Delay Range - 0 ns to
20 s.
Events Delay Range - 2 to 10,000,000
events.
External Rear Input - >1.5 kilohm;
Max input voltage is ±6 V (DC + AC peak).
Video Trigger Types - NTSC, PAL,
SECAM and custom; TV field, field 2 or both, any line within
a field. Line rates -10 kHz to 64 kHz, interlaced, noninterlaced,
composite.
Video Trigger Sensitivity - 0.6
divisions of composite SYNC will achieve a stable display.
Display
Waveform Style - Dots or vectors.
Infinite and variable persistence from 250 ms to 10 s.
Gray Scaling - With variable persistence
selected, waveform points gradually decay through 16 levels
of intensity, providing "z-axis" information about rapidly
changing waveforms.
Update Rate - 200 ea. 500 point
waveforms per sec with infinite persistence mode selected.
Graticules - Full, grid, cross hair,
frame.
Format - YT and XY.
VGA Out - Drives VGA display
monitors.
Zoom
The zoom feature allows waveforms to be
expanded, compressed and positioned in both vertical and horizontal
axes. Allows precise comparison and study of fine waveform detail
without affecting ongoing acquisitions. When used with Hi-res or
Average acquisition modes, Zoom provides an effective vertical dynamic
range of 1000 divisions or 100 screens.
Measurement System
Automatic waveform measurements -
|
|
| Period |
Frequency |
| High |
Low |
| + Width |
- Width |
| Maximum |
Minimum |
| Rise |
Fall |
| Peak to Peak |
Amplitude |
| + Duty cycle |
- Duty cycle |
| + Overshoot |
- Overshoot |
| Propagation delay |
Burst width |
| Mean |
Cycle
mean |
| RMS |
Cycle RMS |
| Area |
Cycle
area |
| Phase |
- |
Continuous update of up to four measurements
on any combination of waveforms.
Thresholds - Settable in percentage
or voltage.
Gated - Any region of the record
may be isolated for measurement using vertical bars.
Snapshot - Performs all measurements
on any one waveform showing results from one instant in time.
Cursor Measurements - Absolute,
Delta, Volts, Time, Frequency.
Cursor Types - Horizontal bars (volts); vertical
bars (time); paired; operated independently or in tracking mode.
Waveform Processing
Waveform Functions - Interpolate-selectable
sin(x)/x or linear, Average, Envelope.
Advanced Waveform Functions - FFT,
Integration, Differentiation (Opt. XL).
Arithmetic Operators - Add, Subtract,
Multiply, Invert.
Autosetup - Single button, automatic
setup on selected input signal for vertical, horizontal and trigger
systems.
Waveform Limit Testing - Compares
incoming waveform to a reference waveform's upper and lower limits.
Computer Interface
GPIB (IEEE 488.2) Programmability -
Full talk/listen modes. Control of all modes, settings and measurements.
Hardcopy/Desktop Publishing
Printer - HP ThinkJet, Epson, PostScript,
Interleaf, DeskJet, LaserJet, TIFF, PCX, BMP (Microsoft Windows).
Plotter - HPGL.
Interface - GPIB standard.
Hardcopy Interface - Centronics
Type and RS-232 (Opt. XL).
Storage
Waveforms - 120,000 waveform points
of nonvolatile storage (with Opt. XL).
Floppy Drive - 3.5 in. 1.44 MB
or 720 KB MS DOS compatible (store waveforms, screen data,
and setups)*1.
Setups - 10 front-panel setups.
CRT
Type - 7 in. diagonal, magnetic
deflection. Horizontal raster-scan. P31 green phosphor.
Resolution - 640 horizontal by 480
vertical displayed pixels.
Power Requirements
Line Voltage Range - 90 to
250 VRMS.
Line Frequency - 48 to 440 Hz.
Power Consumption - 240 W max.
Environmental and Safety
Temperature - Operating: 0°C
to +50°C. Nonoperating: -40°C to +75°C.
Humidity - Operating and nonoperating:
Up to 95% relative humidity at or below +40°C; to 75% relative
humidity from +41°C to +50°C.
Altitude - Operating: 15,000 ft.,
nonoperating: 40,000 ft.
Electromagnetic Compatibility - 89/336/EEC.
Safety - UL3111-1, CSA1010.1, EN61010-1,
IEC61010-1.
Physical Characteristics
| Dimensions |
mm |
in. |
| Height |
164 |
6.4 |
| w/acc. pouch |
177 |
7.5 |
| Width |
362 |
14.25 |
| Depth w/front cover
installed |
491 |
19.33 |
| Depth w/handle extended |
576 |
22.7 |
| Weight |
kg |
lbs. |
| Net approximately |
9.1 |
22.5 |
| Shipping approximately |
12.5 |
32 |
|