TEK TLS216 Logic Scope
Features
The TLS 216 Logic Scope is a new class of instrument designed to simplify
the task of debugging digital hardware. The Logic Scope seamlessly combines
in a single instrument the analog acquisition system of a high-speed digital
storage oscilloscope (DSO) with the triggering and display systems of
a logic analyzer. The 500 MHz bandwidth Logic Scope samples all channels
simultaneously at 2 GS/s and has sophisticated time-qualified triggering,
a high resolution color display and an integrated MS-DOS compatible 3.5
inch floppy disk drive.
Sophisticated Triggering to Identify Complex Digital Problems
In addition to edge trigger, most digital signals can be easily captured
using pulse, glitch and pattern triggers. The Logic Scope provides two
new trigger resources that allow the instrument to directly trigger on
common digital circuit behavior. The industry's first time-interval or
sequence trigger type monitors the time between two events, allowing the
instrument to easily trigger on setup time violations, hold-time violations
or unexpected propagation delay. The powerful "Time-out" trigger
type can be used to capture incomplete handshake sequences or to trigger
the instrument when the DUT "hangs".
All of these trigger types let developers identify channel-to-channel
relationships, including 16-Bit patterns and time-related/time-qualified
system faults. The Logic Scope's external trigger-input can be used as
a "Trigger Arm" to enhance cross-triggering between two instruments,
simplifying the task of using a Logic Scope with other test equipment.
2.5 pF, Low-mass FET Probes Ensure Nonintrusive Connection
The Logic Scope includes a set of 16 specially designed probes that have
extremely low probe-tip mass (1.5 grams) and input capacitance (2.5 pF).
The P6240 low probe-tip mass ensures that connections made to surface
mount and fine pitch ICs will be reliable. The low input capacitance,
combined with the 1 megaohm input resistance, decreases the effect of
the probe on the DUT's operation, allowing very accurate measurements
to be made with confidence.
These characteristics are made possible by using a "podlet-style"
probe-tip design instead of the "pencil-style" design of traditional
oscilloscope probes. Employing the de facto industry standard of 0.1 inch
spacing between the signal and ground inputs, each 0.1 inch thick podlet
can directly attach to the hundreds of readily available IC adapters and
clips.
- Facilitates Hands-free Connection to SMT Devices
- Low-capacitance Probe for Nonintrusive Probing
- Powerful Triggering Reduces Time to Identify Problems
- Flexible Display Modes Simplify Recognition of Timing Faults
- High-speed Acquisitions on All Channels Facilitate Analysis of Complex
System Interactions
Applications
- Hardware Performance Verification
- Multi-channel Data Acquisition
- Mixed Signal Analysis
- A/D and D/A Analysis
- Hardware Timing Analysis
Specifications
| Sample Rate |
2 GS/s |
| Bandwidth |
500 MHz |
| Channels |
16 |
| Samplers |
16 |
| Sensitivity |
50 mV to 10 V/div*1 |
| Position
Range |
±5
Div |
| Offset |
±1 V from 1
to 99.5 mV/div |
Vertical System
DC Gain Accuracy - ±1.5% after
probe calibration.
Vertical Resolution - 8-Bits (256 levels
over 10.24 vertical divisions), 11-Bits with averaging.
Analog Bandwidth Selection - 20 MHz
and full.
Input Coupling - DC.
Input Impedance - 50 Ohm with coax
adapter, 1 megaohm with probe.
Maximum Input Voltage - Max nondestructive
input ±25 V (DC + Peak AC).
Dynamic Range - +15 V to -12 V
with probe.
Time Base System
| Time Bases |
Main,
Delayed |
| Time/Division Range |
500 ps to 5 s/div |
| Time Base
Accuracy |
0.01%
over any interval </=1ms |
| Record Length |
500 to 2,000 pts |
| Pre-trigger
Position |
20%
to 80% of record |
Acquisition Modes
Sample - Sample data only.
Envelope - Max/min values acquired over
one or more acquisitions.
Average - Waveform averages selectable
from 2 to 10,000.
Display
Mixed-mode Data Display - Logic Scope
offers Analog, Timing and BusFormTM display formats that simplify
recognition of digital timing faults and signal anomalies. BusForm display
compresses many input signals into the least amount of display area. The
acquired data can be displayed in any of the three display modes without
re-acquiring.
Waveform Style - Dots, vectors, variable
persistence selectable from 250 ms to 10 s, infinite persistence and intensified
samples.
Color - Standard palettes and user definable
colors for waveforms, text, graticules and cursors. Measurement text and
cursor colors matched to waveform. Waveform collision areas highlighted
with different color. Statistical waveform distribution shown with color
grading through variable persistence.
Color Grading - With variable persistence
selected, historical timing information is represented by temperature
or spectral color scheme providing "z-axis" information about
rapidly changing waveforms.
Graticules - Full, grid, cross hair, frame.
Format - YT.
Fit to Screen - Entire acquisition memory
displayed on screen.
TRIGGERING SYSTEM
Triggers - Main, Delayed.
Main Trigger Modes - Auto, Normal, Single.
Delayed Trigger - Delayed by time and/or
events.
Time Delay Range - 16 ns to 250 s
(time/div </=10 µs); 15.1 ns to 250 s (time/div </=25 µs).
Events Delay Range - 1 to 9,999,999 events.
AUX Trigger Input - TTL compatible; Max
input voltage is -1 V to +6 V.
Measurement System
The Logic Scope provides a comprehensive suite
of automatic measurements specifically designed to improve efficiency
when troubleshooting digital hardware. Twenty-eight on-board measurements,
including setup time, hold time and skew, speed identification of common
digital problems. Whereas traditional DSOs confine waveform measurements
to establishing the relationship of a single channel to a clock, the Logic
Scope simultaneously samples on all channels, so you can see the relationships
between multiple channels.
Automatic Waveform Measurements
| Period |
Frequency |
| High |
Low |
| + Width |
- Width |
| Maximum |
Minimum |
| Rise |
Fall |
| Peak to
Peak |
Amplitude |
| + Duty Cycle |
- Duty Cycle |
| + Overshoot |
-
Overshoot |
| Propagation Delay |
Burst Width |
| Mean |
Cycle
Mean |
| rms |
Cycle rms |
| Area |
Cycle
Area |
| Phase |
Setup Time |
| Hold Time |
Skew |
Continuous update of up to four measurements
on any combination of waveforms.
Gated - Any region of the record may be
isolated for measurement using vertical bars.
Snapshot - Performs all measurements on
any one waveform, showing results from one instant in time.
Cursor Measurements - Absolute, Delta; Volts,
Time, Frequency, Binary Readout.
Cursor Types - Horizontal bars (volts); Vertical
bars (time); operated independently or in tracking mode.
Computer Interface
GPIB (IEEE 488.2) Programmability - Full
talk/listen modes. Control of all modes, settings, and measurements.
Hard Copy/Desktop Publishing
Printer - HP ThinkJet, Epson, PostScript,
DeskJet, LaserJet, DPU 411/412.
Export File Formats - EPS (Encapsulated
PostScript), Interleaf, TIFF, PCX, BMP, RLE.
Plotter Support - HPGL.
I/O Ports - GPIB, Centronics, RS-232 (Talk
only).
Floppy Disk Drive - 3.5 in., 1.44 MB DOS
format.
VGA Output - 15-Pin analog output (Color).
Storage
Waveforms - 16 full 2,000 pt waveforms.
119,008 500 pt waveforms on 1.44 MB floppy disk.
Setups - 10 front-panel setups. 78,947
setups on 1.44 MB floppy disk.
Trigger Types (Main and Delayed)
| Edge |
Trigger when either a rising
or falling edge (positive or negative slope) is detected. |
| Pulse |
The
pulse may be positive or negative and defined by a duration. |
| Width (<Time,
>Time) |
Trigger
when the duration of either a positive or negative pulse is less
than or greater than the user-defined input. |
| Range (In
Range, Out of Range) |
Trigger
when the duration of either a positive or negative pulse is either
between the upper and lower limits or outside the limits. |
| Time-out |
Trigger
at the end of a user-defined time period if the trailing edge of
either a positive or negative pulse is not detected. |
| Pattern |
Each input may be defined
as a low (logic 0), high (logic 1), or don't care (X). Inputs can
be ANDed or ORed. The pattern may be positive true or positive false. |
| Width (<Time, >Time) |
Trigger when the duration of a user-defined
logical pattern is less than or greater than the user-defined input. |
| Range (In Range, Out of Range) |
Trigger when the duration of a user-defined
logical pattern is either between the upper and lower limits or
outside the limits. |
| Time-out |
Trigger at the end of a user-defined
time period if the user-defined logical pattern of a specified duration
does not occur. |
| Sequence |
The
sequence is defined by a set of start and end events. The start
or end event can be either a pattern or an edge (positive or negative
slope). |
| Width (<Time,
>Time) |
Trigger
when the duration between the user-defined start and end events
is less than or greater than the user-defined input. |
| Range (In
Range, Out of Range) |
Trigger
when the duration between the user-defined start and end events
is either between the upper and lower limits or outside the limits. |
| Time-out |
Trigger
at the end of a user-defined time period if the user-defined end
event does not occur within the specified time after the start event. |
| State |
Trigger when a user-specified
logical pattern is detected and the input defined as the clock goes
true or goes false. Each input may be specified as a low (logic
0), high (logic 1), or don't care (X). Inputs can be ANDed or ORed. |
Power Requirements
Line Voltage Range - 90 to 250 V
rms.
Line Frequency - 47 to 63 Hz.
Power Consumption - 300 W max.
Environmental and Safety
Temperature - Operating: +4°C to +50°C.
Nonoperating: -22°C to +60°C.
Humidity - Operating and Nonoperating:
Operating to 80% at or below 29°C, to 20% from +30°C to +50°C.
Nonoperating to 90% at or below 41°C, to 5% from +41°C to 50°C.
Altitude - Operating: 15,000 ft., Nonoperating:
40,000 ft.
Electromagnetic Compatibility - Meets
89/336/EEC.
Safety - UL1244, CSA231, EN61010-1, IEC61010-1.
Probes
Active Probes - 2.5 pF, 1 megaohm FET.
Physical Characteristics
| Dimensions |
mm |
in. |
| Height with feet |
236 |
9.3 |
| without
feet |
193 |
7.6 |
| Width with handle |
445 |
17.5 |
| Depth
with front cover installed |
432 |
17 |
| Weight |
kg |
lbs. |
| Net approximately |
12.3 |
27 |
| Shipping
approximately |
20 |
44 |
|