TEK URV5 Power Meter
Features
The
URV 5 RF Millivoltmeter is a broadband voltage, level and power meter
with high accuracy and sensitivity. It is suitable for both manual operation
and use in systems. A great variety of probes, sensors and accessories
allows the URV 5 to be used for all kinds of measurements.
The
values measured in the two channels can be displayed separately, set off
against one another or referred to any reference. For relative measurements,
the difference, difference in percent, logarithmic difference or ratio
is indicated (DELTAV, DELTAW, DELTA%, DELTAdB, X/REF).
The
URV 5 has a fast trend indication which follows the variations of the
measured value.
URV 35 Level Meter
The
URV 35 Level Meter is a precise voltage and power meter for a large variety
of applications. It can be used in service and production, as well as
for precision measurement in the RF and microwave lab.
The
large variety of probes and sensors available opens up a wide level and
frequency range. Excellent RF shielding provides precise measurements,
even in the near field of antennas.
An
intelligent operating concept in conjunction with an auto-range facility
reduces the user's work to read the results in standard units, either
as absolute or relative values. The appropriate scale is shown on the
LCD display for convenient reading of the results.
Filtering,
which is dependent upon the level and resolution, as well as the choice
between 4 1/2- and 3 1/2-digit readout, optimizes the display of results.
Each probe and sensor is calibrated individually. All relevant data of
the probes are automatically evaluated in the URV 35.
- DC
to 26.5 GHz Frequency Range
- Compact,
Handy and Mobile
- Voltmeter
and Power Meter in One Unit
- Combined
Analog and Digital Display
- Menu-guided
Operation
- DC
Frequency Input for Tracking Frequency-response Correction
- Analog
Output of YT Recorder
- Remote
Control Interface
- Battery
or AC Supply Operation
- Wide
Choice of Probes and Sensors
- Rugged
Design
Applications
- Service
- Production
- Precision
Measurements in RF and Microwave Labs
Specifications
Optional
DC Output (URV 5 with Opt. 01) -
Output voltage range (EMF): -1.999 to +1.999 V into Zout
= 1 kilohm.
Resolution: 1 mV (10 digits).
Error: ±2 mV.
Remote
Control -
Interface: IEC625-1 (IEEE 488) for control of all device functions.
Interface functions: SH1, AH1, T5, L4, SR1, RL1, DC1, DT1, PP1.
Physical Characteristics
| Dimensions |
mm |
in. |
| Height |
110 |
4.3 |
| Width |
241 |
9.5 |
| Depth |
340 |
13.4 |
| Weight |
kg |
lbs. |
| Net |
4.4 |
9.7 |
URV 35
Probes
and Sensors - Compatible with all URV 5-Z probes and NRV-Z sensors.
Frequency
Range - DC to 26.5 GHz.
Display
- Backlit LCD for display of measured value, units and meter scale;
additional moving-coil meter with short response time.
Absolute readout: dBm, dBµV, V or W.
Relative readout: dB referred to stored reference value.
Resolution of digital display -
4 ½ digits: 19,999 steps, 0.001 dB.
3 ½ digits: 1,999 steps, 0.01 dB.
Analog display: Steps of 1/2.5/5 with readout in V, W, or dB;
steps of 5 dB with readout in dBm and windows of 10 dB,
manually or automatically selected; free scaling by entry of left-hand
and right-hand scale limits.
Display filtering: Level-dependent digital averaging filter.
4 ½-digit resolution: Averaging over 16 to 256 readings.
3 ½-digit resolution: Averaging over 1 to 32 readings.
Error
Limits (18 to 28°C) -
Digital display: ±0.02 dB ±1 digit.
Moving-coil meter: 1.5% of full scale.
Zero
Adjustment - Via interface or manually, approximately 4 s.
HOLD
Function - Displayed measurement result retained upon keystroke.
Measurement
Rate - Five readouts per second in manual mode.
Frequency-response
Correction - Specific calibration data of probe taken into account;
frequency entry via keypad, interface or equivalent DC voltage at rear-panel
input.
Attenuation
Correction - Attenuation or gain connected to input taken into account;
input of attenuation value via serial interface or keypad, range ±199.99 dB.
Reference Value
Entry - Measured value on keystroke or value entered via interface
or keypad.
Reference
Impedance - 50 Ohm/75 Ohm, automatic/selectable.
Input/Output
-
DC frequency input:
Connector: BNC female.
Input impedance: 9 megaohm.
Input voltage range: ±12 V, freely selectable, linear, scaling.
Maximum input voltage: 50 V.
DC Voltage Output:
Connector: BNC female.
Source impedance: 1 kilohm.
Output voltage (EMF): Proportional to pointer deflection.
Left-hand scale limit: Corresponding to 0 V.
Right-hand scale limit: Corresponding to 0 V.
Settling time: 250 ms.
Error: </=5 mV.
Ripple: 5 mV p-p typical.
Remote
Control - All device functions controlled via serial interface (V.24,
RS-232C).
Power
Requirements (URV 5 and URV 35)
Battery
Operation (URV 35 Model 02) -
Battery type: 5 LR20 (1.5 V).
Operating time: Approximately 125 hours.
Charge time (with UZ-35 charger): Approximately 24 hours.
AC
Operation (URV 5 and URV 35 Model 03) -
Line voltage: 100/120/220/240 VAC ±10%.
Line frequency: 47 to 63 Hz, 400 Hz.
Power consumption: 30 VA.
Environmental
Characteristics (URV 5 and URV 35)
Temperature
Range -
Operating: 0 to +50°C (Class I to IEC 359, no condensation).
Storage: -40 to +70°C.
Humidity
- 80% maximum (without condensation).
Mechanical
Strength - To IEC 359, Class I.
RFI
Suppression - VDE 0871, DBP approval number to regulations 527/1979.
EMC
- To IEC 801.3, severity level 3; MIL-STD-461-B RS 03 and CS 02.
Physical Characteristics
| Dimensions |
mm |
in. |
| Height |
103 |
4.1 |
| Width |
219 |
6.6 |
| Depth |
240 |
9.4 |
| Weight |
kg |
lbs. |
| Model
02 |
|
|
| With
batteries |
3.1 |
6.8 |
| Without
batteries |
2.3 |
5.1 |
| Model
03 |
2.4 |
5.3 |
|