P6810
|
P6860
|
P6880
|
P6417
|
P6418
|
P6434
|
|
Logic Analyzer Used
|
TLA7AAx Logic Analyzer Modules
TLA7ABx Logic Analyzer Modules
|
TLA6xx Logic Analyzers
TLA7Lx/7Mx Logic Analyzer Modules
TLA7Nx/7Px/7Qx Logic Analyzer Modules
|
|
Recommended Use
|
Recommended for most general-purpose uses that require maximum flexibility
for single-ended or differential requirements
|
Recommended for applications requiring many channels to be quickly connected
in a small footprint
|
Recommended for applications requiring many differential channels to
be quickly connected in a small footprint
|
Recommended for most general-purpose uses that require maximum flexibility
|
Recommended for most general-purpose uses
|
Recommended for applications requiring many channels to be quickly connected
in a small footprint
|
|
Attachment to Target System
|
Probe leadsets adapt to industry standard interfaces; leads spread over
a wide area
|
Connectorless "compression" contact (Adapter for Mictor connector available)
|
Connectorless "compression" contact (Adapter for Mictor connector available)
|
Probe leadsets adapt to industry standard interfaces; leads spread over
a wide area
|
Probe leadsets adapt to industry standard interfaces
|
AMP Mictor 34 channel connector (Adapter to use P6434 with P6860/80 high-density
compression land footprint available)
|
|
Probe Type
|
General purpose, 34 channel active probe
|
High density, 34 channel active probe
|
High density, 34 channel active differential probe
|
General purpose, 17 channel passive probe
|
General purpose, 17 channel passive probe
|
High density, 34 channel passive probe AMP Mictor connector required
|
|
Pin Spacing Supported
|
0.100 in. and 2 mm
|
N/A
|
N/A
|
0.100 in.
|
0.100 in.
|
N/A
|
|
Logic Signals Supported
|
Differential Clock Differential Data
|
Differential Clock Single-ended Data
|
Differential Clock Differential Data
|
Single-ended Clock and Data (Differential signal adapters available)
|
Single-ended Clock and Data (Differential signal adapters available)
|
Single-ended Clock and Data
|
|
Simultaneous State/Timing to:
|
800 MHz/ 8 GHz
|
800 MHz/ 8 GHz
|
800 MHz/ 8 GHz
|
200 MHz/ 2 GHz
|
200 MHz/ 2 GHz
|
200 MHz/ 2 GHz
|
|
Simultaneous State/Timing/ Analog to:
|
800 MHz/ 8 GHz/ 2 GHz
|
800 MHz/ 8 GHz/ 2 GHz
|
800 MHz/ 8 GHz/ 2 GHz
|
N/A
|
N/A
|
N/A
|
|
Minimum Signal Amplitude Minimum Single-ended
|
300 mVp-p
|
300 mVp-p
|
300 mVp-p
|
500 mVp-p
|
500 mVp-p
|
500 mVp-p
|
|
Minimum Differential
|
Vmax - Vmin >150 mV
|
Vmax - Vmin >150 mV
|
Vmax - Vmin >150 mV
|
N/A
|
N/A
|
N/A
|
|
Probe Load AC/DC
|
1.2 pF/20 kohm to Ground
|
0.7 pF/20 kohm to Ground
|
0.7 pF/20 kohm to Ground
|
2 pF/20 kohm to 2.2 V (Low-voltage adapters that work
with low-voltage signals are available)
|
2 pF/20 kohm to 2.2 V (Low-voltage adapters that work
with low-voltage signals are available)
|
2 pF/20 kohm to 2.2 V
|
|
Notes
|
Works with a wide-range of industry-standard accessories for flexible
attachment to your target system
|
No connector required - only land pads required to be laid out on target
system PCB for 17 and/or 34 channels. Please refer to P6860/6880 probe
design guide
|
No connector required - only land pads required to be laid out on target
system PCB for 17 and/or 34 channels. Please refer to P6860/6880 probe
design guide
|
Works with a wide-range of industry-standard accessories for flexible
attachment to your target system
|
Works with a wide-range of industry-standard accessories for flexible
attachment to your target system
|
Requires AMP Mictor connector to be installed on target system PCB for
every 34 channels.Please refer to P6434 probe manual
|